Infrared Controls (for Infrared Self-Activated Sinks only)
Installation
The sinks are supplied with a 24V power transformer(s) that connects to a standard duplex outlet (110/120V
outlet required).
Single basin sinks have one sensor, dual basin sinks have two sensors (one for each basin) and triple basin sinks
have three sensors (one for each basin).
1. Plug the transformer(s) into the outlet.
2.
A red LED will flash in the sensor window (Fig. 16 - circled in red).
A.
Important:
Do not interrupt the sensor beam until the light turns off.
Operation
The sensors are pre-set and equipped with a logic board. The sensors determines the range during initialization
period (The time after initial power until the light turns off is approximately 5 minutes). The range is approximately
12-14” in front of the sensor and is 25 degrees at peak.
During the initialization period, the sensors allow for fixed objects that may be within the sensors’ range. The
sensors are equipped with a 2 second on/off delay, and no-time-out feature. This prevents the sink from turning
on when walking past at a normal pace and no-time-out allows for an uninterrupted scrub.
Fig. 16: Infrared sensor on sink front