1
INSTALLATION INSTRUCTIONS:
Replacing a Standard Warmer
Header (DWC24-TL) with a Data Log Header (DWC24-TL-D)
IFU - 033 Rev.A
02/23/2015
A: Removing Standard Warmer Cabinet Header
(p/n: DWC24-TL)
The Standard Warmer Header for a DWC24-TL
Dual Warmer Cabinet can be easily replaced
with a DWC24-TD-D Warmer Cabinet Header
with a Data Logger option.
1. Disconnect all electrical power to the Warming
Cabinet by removing the power cord from the
electrical supply. If the unit is hard-wired to
facility power, turn
OFF
the circuit breaker.
CAUTION:
Failure to remove the power cord from
the electrical supply can result in severe
electrical shock and even death to
personnel as well as severe damage to
the equipment.
2. The Outside and Inside Top Panels must be
removed first. See Figure 1.
3. Remove four 8 X 1-5/8” self-tapping screws
from Outside Top Panel. Remove Outside
Top Panel and set aside. See Figure 1.
4. Remove two 8 X 1/2” self-tapping screws from
Inside Top Panel. Remove Inside Top Panel
and set aside. See Figure 1.
The Standard Warmer Header for a DWC24-TL
Dual Warmer Cabinet can be removed from a
standard dual warming cabinet without removing
the Upper Drawer Assembly (p/n: W0288).
1. There is no specific order in removing all
wiring to the Header, just specific steps that
allows the wiring to pass safely through the
electrical box.
2. Loosen two screws on the 3/8” straight-
through electrical connector mounted on the
back side of the Warmer Cabinet Electrical
Box (just behind the Header). See Figure 2.
3. Use a Pin Extraction Tool (for 16AWG wire) to
remove the wire leads from the 9-/5-Connector
Circuit Plugs. See Figure 3.
CAUTION: Use ONLY a Pin Extraction Tool
designed for the specific wire size
REF
P/N
DESCRIPTION
QTY
1
SMW0027
Outside Top Panel
1
2
SMW0028
Inside Top Panel
1
3
H0012-01
Screw, Self-Tapping, 8 X 1/2”
2
4
H0012-02
Screw, Self-Tapping, 8 X 1-5/8”
4
Figure 1: Removing Inside and Outside Top Panel
1
2 3
4
REF
P/N
DESCRIPTION
QTY
1
W0199
Connector, 9 Circuit Plug
1
2
W0198
Connector, 5 Circuit Plug
1
3
SMW0032
Cover, Electrical Box
1
4
W0140
Connector, 3/8” Straight
1
REF
P/N
DESCRIPTION
QTY
1
16AWG
Tool, Pin Extraction
1
2
W0198
Connector, 5 Circuit Plug
1
3
16AWG
Wire, Crimped Terminal End
1
Figure 2: Loosening Wire Connector on Electrical Box
1
2
3
4
Figure 3: Removing Wires from Circuit Plug
1
2
3