INSTALLATION INSTRUCTIONS: 

  Replacing a Standard Warmer

Header (DWC24-TL) with a Data Log Header (DWC24-TL-D)

IFU - 033 Rev.A

02/23/2015

A:  Removing Standard Warmer Cabinet Header 

(p/n: DWC24-TL)

The Standard Warmer Header for a DWC24-TL 
Dual Warmer Cabinet can be easily replaced 
with a DWC24-TD-D Warmer Cabinet Header 
with a Data Logger option.
1. Disconnect all electrical power to the Warming 

Cabinet by removing the power cord from the 
electrical supply. If the unit is hard-wired to 
facility power, turn 

OFF

 the circuit breaker.

CAUTION:  

Failure to remove the power cord from 

the electrical supply can result in severe 
electrical shock and even death to 
personnel as well as severe damage to 
the equipment.

2. The Outside and Inside Top Panels must be 

removed first. See Figure 1.

3. Remove four 8 X 1-5/8” self-tapping screws 

from Outside Top Panel. Remove Outside 

Top Panel and set aside. See Figure 1. 

4. Remove two 8 X 1/2” self-tapping screws from 

Inside Top Panel. Remove Inside Top Panel 

and set aside. See Figure 1.

The Standard Warmer Header for a DWC24-TL 
Dual Warmer Cabinet can be removed from a 
standard dual warming cabinet without removing 
the Upper Drawer Assembly (p/n: W0288).

1. There is no specific order in removing all 

wiring to the Header, just specific steps that 
allows the wiring to pass safely through the 
electrical box.

2. Loosen two screws on the 3/8” straight- 

through electrical connector mounted on the 

back side of the Warmer Cabinet Electrical 

Box (just behind the Header). See Figure 2.

3. Use a Pin Extraction Tool (for 16AWG wire) to 

remove the wire leads from the 9-/5-Connector 

Circuit Plugs. See Figure 3.

CAUTION:  Use ONLY a Pin Extraction Tool 

designed for the specific wire size 

REF

P/N

DESCRIPTION

QTY

1

SMW0027

Outside Top Panel

1

2

SMW0028

Inside Top Panel

1

3

H0012-01

Screw, Self-Tapping, 8 X 1/2”

2

4

H0012-02

Screw, Self-Tapping, 8 X 1-5/8”

4

Figure 1:  Removing Inside and Outside Top Panel

1

2 3

4

REF

P/N

DESCRIPTION

QTY

1

W0199

Connector, 9 Circuit Plug

1

2

W0198

Connector, 5 Circuit Plug

1

3

SMW0032

Cover, Electrical Box

1

4

W0140

Connector, 3/8” Straight

1

REF

P/N

DESCRIPTION

QTY

1

16AWG

Tool, Pin Extraction

1

2

W0198

Connector, 5 Circuit Plug

1

3

16AWG

Wire, Crimped Terminal End

1

Figure 2:  Loosening Wire Connector on Electrical Box

1

2

3

4

Figure 3:  Removing Wires from Circuit Plug

1

2

3

Summary of Contents for DWC24-TL

Page 1: ...al Warmer Cabinet can be removed from a standard dual warming cabinet without removing the Upper Drawer Assembly p n W0288 1 There is no specific order in removing all wiring to the Header just specific steps that allows the wiring to pass safely through the electrical box 2 Loosen two screws on the 3 8 straight through electrical connector mounted on the back side of the Warmer Cabinet Electrical...

Page 2: ...ed to be taken apart Item number 2 is a female receptacle p n W0038 and must have the wires removed before the header can be removed 7 Remove the two screws on one female receptacle and one male receptacle and lift off the cover See Figure 5 Items 2 and 3 Place the screws and cover in an area where they will not be misplaced Loosen the two screws on the inside and pull the two wires away from the ...

Page 3: ...e black and one white into the Lower Cabinet 9 Position Molded Plug See Figure 9 Item 1 Reconnect the two halves of the plug Make sure the wire colors match once the plug is reconnected 4 Insert the barbed connector wires one black and one red into the Upper Cabinet Connector REF DESCRIPTION QTY 1 Power Cable Ground Wires W 1 2 To Lower Chamber Control Board W 1 3 To Upper Chamber Control Board W ...

Page 4: ...two bundles are connected separate the wires into two groups One group goes to the Upper Cabinet and the other group goes to the Lower Cabinet Use a cable tie to secure and separate each of the two bundles Re install the Outside and Inside Top Covers 1 Install the Inside Top Panel using two 8 X 1 2 inch screws to the cabinet See Figure 12 Item 4 2 Use four 8 X 1 5 8 self tapping screws to secure t...

Reviews: