Blanket Warmer Operating Manual 

 

   

 

 

 

 

 

 

 

 
 

 

 

Summary of Contents for 3MW1-030004

Page 1: ...Distributed by Owens Minor Mechanicsville VA 23116 For Blanket Warmer Service call 800 860 4716 Blanket Warmer Operating Manual ...

Page 2: ...inets 4 Half Size Blanket Warming Cabinets 5 Counter Top Blanket Warming Cabinets 6 GENERAL INFORMATION 7 Start Up Instructions 8 Display Instructions 9 Technical Data on Digital Controller 10 Alarm Settings 11 Maintenance 12 CUVERRO ANTI MICROBIAL HANDLES 13 TEMPERATURE RECOMMENDATIONS 14 BLANKET RECOMMENDATIONS 15 PARTS LIST 16 LIMITED WARRANTY 17 ...

Page 3: ...zed for left side with an up charge Low Voltage digital controls are programmable and lockable with LED illumination preset visual alarms and preset temperature for each compartment and are located on the front top service compartment Temperature may be set between 60 and 140 degrees F plus or minus 2 degrees F The visual alarm may be set at plus or minus 1 degree F with the temperature at 100 deg...

Page 4: ...ro Anti microbial Handle v Electrical is 120V 50 60 Cycles Highly visible illuminated on off switch located at top of service compartment Energy efficient commercial grade fans circulate heated air thru vented stainless steel ducts v Low voltage digital controls are programmable and lockable with LED screen preset alarms and preset temperature Temperature may be set between 60º and 140º F 2º F Ala...

Page 5: ... x 30 H v CuVerro Anti microbial Handles v Electrical is 120V 50 60 Cycles Highly visible illuminated on off switch located at top of service compartment Energy efficient commercial grade fans circulate heated air thru vented stainless steel ducts v Low voltage digital controls are programmable and lockable with LED screen preset alarms and preset temperature Temperature may be set between 60º and...

Page 6: ...v CuVerro Anti microbial Handles v Electrical is 120V 50 60 Cycles Highly visible illuminated on off switch located at top of service compartment Energy efficient commercial grade fans circulate heated air thru vented stainless steel ducts v Low voltage digital controls are programmable and lockable with LED screen preset alarms and preset temperature Temperature may be set between 60º and 140º F ...

Page 7: ...Anti microbial Handles v Electrical is 120V 50 60 Cycles Highly visible illuminated on off switch located at top of service compartment Energy efficient commercial grade fans circulate heated air thru vented stainless steel ducts v Low voltage digital controls are programmable and password lockable with LED screen preset alarms and preset temperature Temperature may be set between 60º and 140º F 2...

Page 8: ...ase 3 Wire 5 0 Amps v Counter Top 120 V 50 60 Cycle single phase 3 Wire 2 9 Amps Do not expose to water or moisture Use the controller only within the operating limits avoiding sudden temperature changes with high atmospheric humidity to prevent formation of condensation Disconnect all electrical connections before any kind of maintenance The digital controller Model XR01CX should only be serviced...

Page 9: ...of 1 hour to heat You will know it has reached the pre set temperature of 130 F when LED goes off On initial start up temperature alarm is delayed for 120 minutes to allow temperature to set Your warmer has been set to 130 F Temperature will vary 2 degrees The Alarm is set at a Minimum 90 F and a Maximum 150 F Front Panel Display ...

Page 10: ...erature will be displayed and the C or F LED starts blinking 3 You can now raise or lower temperature with or keys 4 After you set your desired temperature push the SET key again or wait 10 seconds Note You cannot increase the set point temperature higher than the maximum temperature it was originally set to with the above procedure 150 F HOW TO LOCK UNLOCK THE KEYBOARD To LOCK the keyboard press ...

Page 11: ... 50 60 Hz Power Absorption 3 5 VA max Display 2 digits red LED 14 2 mm high Inputs 1 NTC Digital Input Free voltage contact Relay Outputs SPST 20 8 A 250VAc Data Storing On the non volatile memory EEPROM Kind of Action 1B Pollution Grade 2 Rated Impulsive Voltage 2500V Overvoltage Category II Operating Temperature 0 60 C Storage Temperature 25 60 C Relative Humidity 20 85 no condensing Measuring R...

Page 12: ... always active HA Maximum Temperature Alarm Outputs Unchanged LA Minimum Temperature Alarm Outputs Unchanged EA External Alarm Outputs Unchanged CA Serious External Alarm ALL Outputs OFF dA Door Open Unit and fans restart ALARM RECOVERY Probe alarm P1 starts seconds after the fault in the related probe the alarm automatically stops after the probe restarts the normal operation Check connections be...

Page 13: ...very poor quality and produce large amounts of lint that may become attached to the air channel walls See recommendation on blankets Maintenance Procedure 1 Disconnect blanket warmer from power 2 Open blanket warmer door 3 Remove blankets from racks and remove racks from unit 4 Remove back wall retaining screws using 5 16 socket 5 Vacuum out excess lint behind back wall including areas around the ...

Page 14: ... to and not a substitute for standard infection control practices Users must continue to follow all current infection control practices including those practices related to cleaning and disinfection of environmental surfaces This surface has been shown to reduce microbial contamination but does not necessarily prevent cross contamination This surface may be subject to recontamination and the level...

Page 15: ...be consistent ECRI and AORN recommend that the fluid warmer temperature and the length of time that fluids should remain in the warming cabinet should be requested from the manufacturer of each fluid or solution since the requirements may be different for each As general guidance the AORN Recommendation for the Practice Environment of Care Recommendation VIII states that fluids and solutions shoul...

Page 16: ...ncrease in temperatures above 110F creates a potentially unsafe and dangerous environment for some patients We Do Recommend Knitted Warming Blankets 1 Knitted blankets made of a cotton polyester blend 2 They do not shrink or deteriorate as rapidly Usually last 5 to 6 times longer than cheap woven imports 3 They are thermal and have a far superior insulating capability and will sustain a patient wa...

Page 17: ...PARTMENT BLANKET WARMER DOOR NON LATEX DARK CHARCOAL GRAY FITS GLASS OR SOLID DOOR 3MW1 070805 MAGNETIC GASKET FOR TOP DOOR OF DOUBLE DOOR WARMING CABINET NON LATEX DARK CHARCOAL GRAY ALSO FITS UNDER COUNTER HALF SIZE CABINETS 3MW1 070818 MAGNETIC GASKET FOR LOWER DOOR OF DOUBLE DOOR WARMING CABINET NON LATEX DARK CHARCOAL GRAY 3MW1 080163 CHR HEATER BAR 120 VOLT 250 WATT 3MW1 190174 ILLUMINATED O...

Page 18: ...r shall not be liable for the following v Repairs and replacements because of misuse abuse negligence alteration accident freight damage or tampering v Products considered to be of a consumable nature v Accessories or parts not manufactured by the Manufacturer v Charges by anyone for adjustments repairs replacement parts installation or other work performed upon or in connection with such products...