EM-1712  Rev.4  P. 2 / 4

FRONT VIEW & PROGRAMMING

 

PROGRAMMING PROCEDURE

1) Press ITEM UP or DOWN key until ITEM display indicates “01”.
2) Press DATA UP or DOWN key and choose “2” on DATA display.

1: Data indication only.
2: All parameters are modifiable.

3) Press ITEM UP or DOWN key until ITEM display shows the 

ITEM No. you need to change.

4) Press DATA UP or DOWN key and choose a DATA No. or value 

you need on DATA display.

5) Repeat above 3 and 4.  (Entered data is stored 1 sec. after the 

operation has been complete.)

6) Press ITEM UP or DOWN key until ITEM display indicates “01”.
7) Press DATA UP or DOWN key and choose “1” on the display.
8) Press ITEM UP or DOWN key until ITEM display indicates “P”.  

DATA display shows process input.  You can now check data set-
ting by choosing ITEM No.

Note : DO NOT press UP and DOWN keys simultaneously.

ITEM

MDF.

CODE

DATA

CONTENTS

DEFAULT

P

N/A

-9999 – 9999

Input display in engineering unit *

1

---

01

1, 2, 3

Modification code

1 : Data indication only.
2 : All parameters are modifiable.
3 : Only ITEM 19 is modifiable.

1

02

N/A

0 – 99

Status indication (“0” is normally indicated.)

---

03

2

0 – 12

Thermocouple type
0: (PR), 1: K (CA), 2: E (CRC), 3: J (IC), 4: T (CC), 5: B (RH),
6: R, 7: S, 8: C (WRe 5-26), 9: N, 10: U, 11: L, 12: P (Platinel II)

1: K (CA)

04

N/A

0, 1, 2

Output range code

0 : V1 (-1 – +1V)
1 : V2 (-10 – +10V)
2 : Z1 (0 – 20mA)

V1 : 0
V2 : 1
Z1 : 2

05

N/A

-15.0 – 115.0

Input indicated in % (of the range set in ITEM 13/14)

---

06/L

2

-15.0 – 115.0

Output indicated in % with ITEM 01 DATA 1 (of the range set in ITEM 15/16)
Loop test output with ITEM 01 DATA 2 (‘L’ is indicated as ITEM No.)
(Use DATA UP/DOWN key to set the output signal.)

---

07

2

-99.99 – 99.99

Zero adjustment (%) (fine adj. of the value set in ITEM 15)

0.00

08

2

0.000 – 9.999

Gain adjustment (fine adj. of the value set in ITEM 16)

1.000

09

2

0, 1, 2

Temperature unit 

0: °C 

1: °F 

2: K

0

10

2

0, 1, 2, 3, 4

Moving average (sampling cycle: 10 msec.)
0: No 

1: 4 samples 

2: 8 samples 

3: 16 samples 

4: 32 samples

0

11

2

0, 1 – 60

Power-saving mode

0 : Continuous display
1 – 60 : Time before display turned off (minutes)

10

12

2

0, 1

Burnout protection

0 : Downscale
1 : Upscale

1

13
14

2
2

-270.0 – 2410
-270.0 – 2410

Lower range temperature (ITEM 13 < ITEM 14) *

1

Upper range temperature (ITEM 13 < ITEM 14) *

1

0.0
100.0

15
16

2
2

-1.00 – 1.00
-1.00 – 1.00

Output code V1

0% output voltage (V) *

2

100% output voltage (V) *

2

-1.00
1.00

15
16

2
2

-10.0 – 10.0
-10.0 – 10.0

Output code V2

0% output voltage (V) *

2

100% output voltage (V) *

2

-10.0
10.0

15
16

2
2

0.0 – 20.0
0.0 – 20.0

Output code Z1

0% output current (mA) *

2

100% output current (mA) *

2

4.0
20.0

17

2

0, 1

Output characteristic   0: Positive 

1: Negative (inverted output)

0

18

3

0, 1

Reset all settings *

3

0

19

N/A

---

ROM version

---

*1 :  According to the temperature unit specified in ITEM 09.  For under -1000 or over 1000 values, under decimal point position cannot 

be displayed and setting is unavailable.

*2 : ITEM 15 < ITEM 16.
*3 :  Press DATA UP key and choose DATA 1.  Double-click DATA DOWN key.  The display shows DATA 0 after the initialization is 

complete.

PL4

PL3

PL2

PL1

DATA

ITEM

DATA

ITEM

ITEM No. Display

DATA UP-DOWN Keys

ITEM No. UP-DOWN Keys

DATA Display

Status LED

Polarity LED