background image

MLS

P. 3 / 3

EM-1644

CHECKING

1) Terminal wiring:  Check that all cables are correctly

connected according to the connection diagram.

2) Power input voltage:  Check voltage across the terminal 7

–␣ 8 with a multimeter.

3) Inputs:  Check that the input signals are within 0 – 100%

of the full-scale.

4) Output:  Check that the load resistance meets the de-

scribed specifications.

ADJUSTMENT PROCEDURE

This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any calibra-
tion.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.

■ 

HOW TO CALIBRATE THE OUTPUT SIGNAL

Use a signal source and measuring instruments of sufficient
accuracy level.  Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO:  Apply the input signals which should lead the

calculated result as 0%, and adjust output to 0%.

2) SPAN:

[K

1

 

×

␣ K

2

 

 1] Apply the input signals which should lead the

calculated result as 100%, and adjust output to 100%.
[K

1

 

×

␣ K

2

 < 1] Apply the input signals which should lead the

calculated result as [100 

×

 K

1

 

×

␣ K

2

] %, and adjust output to

[100 

×

 K

1

 

×

␣ K

2

] %.

3) Check ZERO adjustment again.
4) When ZERO value is changed, repeat the above procedure

1) – 3).

EQUATION

Output = K

1

 

×

 Input 1 

×

␣ K

2

 

×

␣ Input 2

K

1

, K

2

:

0.10 –␣ 1.15 (parameters)

Input 1, Input 2: 0 –␣ 115%;

Forcibly limited to 0% or 115% with overrange.

Output: 0 – 115% with K

1

 

×

 K

2

 

 1;

Forcibly limited to 0% or 115% with overrange.
0 – [115 

×

␣ K

1

 

×

 K

2

] % with K

1

 

×

 K

2

 < 1;

Forcibly limited to 0% or [115 

×

␣ K

1

 

×

 K

2

] %

with overrange.

K

1

, K

2

 are ex-factory specified.

[example] K

1

K

2

INP1

INP2

OUT

1.00

1.15 115%

100%

115%

1.00

1.15 100%

  50%      57.5%

1.00

1.15 100%

   -5%

    0%

POINTS OF CAUTION

■ 

GENERAL

•Before you remove the unit from its base socket or mount it,
turn off the power supply and input signal for safety.

■ 

ENVIRONMENT

• When heavy dust or metal particles are present in the air,
install the unit inside proper housing and ventilate it.
• Do not install the unit where it is subjected to continuous
vibration.  Do not apply physical impact to the unit.
• Environmental temperature must be within -5 to +60

°

C (23

to 140

°

F) with relative humidity within 30 to 90% RH in order

to ensure adequate life span and operation.
• Be sure that the ventilation slits are not covered with
cables, etc.

■ 

WIRING

• Do not install cables (power supply, input and output) close
to noise sources (relay drive cable, high frequency line, etc.).
• Do not bind these cables together with those in which noises
are present.  Do not install them in the same duct.

■ 

AND ....

• The unit is designed to function as soon as power is
supplied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data sheet.

MAINTENANCE

Regular calibration procedure is explained below:

■ 

CALIBRATION

[K

1

 

×

␣ K

2

 

 1]

Warm up the unit for at least 10 minutes.  Apply the input
signals which should lead the calculated results as 0%, 25%,
50%, 75% and 100%.  Check that the output signal for the
respective input signal remains within accuracy described in
the data sheet.
[K

1

 

×

␣ K

2

 < 1]

Warm up the unit for at least 10 minutes.  Apply the input
signals which should lead the calculated result as 0%, [25 

×

K

1

 

×

␣ K

2

] %, [50 

×

 K

1

 

×

␣ K

2

] %, [75 

×

 K

1

 

×

␣ K

2

] %, and [100 

×

 K

1

×

␣ K

2

] %.  Check that the output signal for the respective input

signals shows 0%, [25 

×

 K

1

 

×

␣ K

2

] %, [50 

×

 K

1

 

×

␣ K

2

] %, [75 

×

 K

1

×

␣ K

2

] %,  and [100 

×

 K

1

 

×

␣ K

2

] %, and it remains within accuracy

described in the data sheet.

When the output is out of tolerance, recalibrate the unit
according to the "ADJUSTMENT PROCEDURE" explained
earlier.

M-SYSTEM WARRANTY

M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System's sole liability, and purchaser's exclusive remedies,
under this warranty are, at M-System's option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this warranty.  To
submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.  M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: [email protected]

Reviews: