background image

 

MB-08, MB-16

1) Install the MB onto a panel surface with the attached 2 

pan-head screws (M5 × 40).

2) Remove the closing panels on which the Transducers are 

to be mounted.  Be sure to keep the gasket attached to the 

closing panel so that it will be reused when attaching the 

MB to the base socket.

  For the P-UNIT Transducers, one unit requires a space 

for two closing panels.  Remove only the right closing 

panel, seen from the front when the pressure supply con-

nections to the MB is positioned at the bottom.

3) Loosen the two screws (M3 × 16) attaching the Trans-

ducer body to its base socket to separate them.

4) Mount the base socket onto the MB with the two screws 

(M4 × 20) provided with the Transducer.  Be sure to in-

sert the gasket between them.

5) Attach the pressure supply pipe referring to ‘Piping.’
6) Plug the Transducer body into its base socket and fasten 

the M3 screws which had been loosened in the above pro-

cedure (3).

Note 1:  Put the closing panel back and the gasket onto the MB 

if the base socket is not attached immediately.

Note 2:  Remove the closing panels at the last moment before 

mounting the base socket, in order to prevent any dust 
entering into the MB.

Note 3:  When installing an additional base socket to an al-

ready working MB, stop the air supply.

EXTERNAL DIMENSIONS  

unit: mm (inch)

2–4.3 (.17) dia.

MTG HOLE

MB-01

75 (2.95)

approx.

27

(1.06)

10 (.39)

1/4” NPT FITTING

Rc 1/4

152 (5.98)

215 (8.46)

225 (8.86)

MB-08

178 (7.01)

418 (16.46)

428 (16.85)

2–3/8” NPT FITTING

2–Rc 3/8

4–6 (.24) dia.

MTG HOLE

3–R3/8 PLUG

3–R3/8 PLUG

CLOSING PANEL

14 (.55)

35 (1.38)

27 (1.06)

36 (1.42)

178 (7.01)

MB-16

75 (2.95)

approx

.

31

(1.22)

54 (2.13)

3/8” NPT FITTING

Rc 3/8

4–6 (.24) dia.

MTG HOLE

CLOSING PANEL

14 (.55)

35 (1.38)

27 (1.06)

36.5

(1.44)

75 (2.95

)

approx

.

31

(1.22)

54 (2.13)

13.5 (.53)

CLOSING PANEL

14 (.55)

35 (1.38)

27 (1.06)

25 (.97)

15 (.59)

2–5 (.20) dia. MTG HOLE
COUNTERBORE,

8 (.31) dia., 10 (.39)deep

MB-P1

75 (2.95)

approx

.

27

(1.06)

10 (.39

)

1/4” NPT FITTING

Rc 1/4

CLOSING PANEL

14 (.55)

35 (1.38)

27 (1.06)

50.5 (1.99)

40.4 (1.59)

MB

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-3430  Rev.2  P. 4 / 5