SETTING UP THE RECORDER
Basic setup procedure for 71VR1-E101-x is explained in this
section, using an example of voltage input measurement at
Input 1 displaying 0 to 100 % to store data in a memory card
in 500 m sec. sample rate.
■
SYSTEM SETTING
1) Hold down F1 button on TREND display screen, DIGIT
-
AL display screen or CONTROL MENU until SETTING
MENU is displayed.
2) SYSTEM is selected. Press
[
ENTER
]
(F6) button to dis-
play SYSTEM SETTING menu.
3) Confirm the settings such as DATE, TIME, NUMBER
OF DECIMALS, and MODBUS. If you need to change
them, press F1 or F2 button to move to the setting you
need to change.
4) Press
[
EDIT
]
(F6) button to change color of the value to
magenta, then change the value by pressing F2 through
F5 button. After changing the value, press
[
ENTER
]
(F6) to set the value.
5) Press
[
RETURN
]
(F1) button to return to TREND dis-
play screen.
■
INPUT CHANNEL SETTING
1) Hold down F1 button on TREND display screen, DIGIT
-
AL display screen or CONTROL MENU until SETTING
MENU is displayed.
2) Press F2 or F3 button to select CHANNEL in the menu
list, then press
[
ENTER
]
(F6) button to display CHAN
-
NEL MENU.
3) OPERATING MODE is selected. Press
[
ENTER
]
(F6)
button to display CHANNEL/OPERATING MODE
menu.
4) SAMPLE RATE is selected. Press
[
EDIT
]
(F6) button.
Press F2 or F3 button to select 500 ms of sample rate.
Press
[
ENTER
]
(F6) button to set the sample rate.
5) NORMAL/DEMO is for selecting NORMAL mode or
DEMO mode. Confirm it is NORMAL.
6) Press
[
RETURN
]
(F1) button to return to the CHANNEL
MENU.
■
CHANNEL/ANALOG INPUT/CHANNEL SETTING
• CHANNEL/ANALOG INPUT menu 1/3
1) Press F2 or F3 button in CHANNEL MENU to select
ANALOG INPUT, then press
[
ENTER
]
(F6) button to
display CHANNEL/ANALOG INPUT menu.
2) Press
[
EDIT –> –>
]
(F6) button to select “AI 1” for INPUT
CHANNEL.
3) Press F2 or F3 button to select FIELD CHANNEL, then
press
[
EDIT –> –>
]
(F6) button to enable editing.
4) Press F2 or F3 button to select Ai1, then press
[
ENTER
]
(F6) button to set the field channel to Ai1.
5) Press F2 or F3 button to select TYPE, then press
[
EDIT
–> –>
]
(F6) button to enable editing.
6) Press F2 or F3 button to select “-10 – 10V”, then press
[
ENTER
]
(F6) button to set the type to “-10 – 10V”.
7) Press F2 or F3 button to select MIN, then press
[
EDIT –>
–>
]
(F6) button to enable editing.
8) Press F2 through F5 button to change the value to 0.000,
then press
[
ENTER
]
(F6) button. The number of decimal
places displayed here is fixed by the setting in the SYS-
TEM SETTING menu.
9) Press F2 or F3 button to select MAX, then press
[
EDIT
–> –>
]
(F6) button to enable editing.
10)Press F2 through F5 button to change the value to
10.000, then press
[
ENTER
]
(F6) button.
• CHANNEL/ANALOG INPUT menu 2/3
To change ENGINEERING UNIT and TAG NAME, the
Configurator Software (model: 71VRCFG) is required.
1) Press F4 or F5 button to display CHANNEL/ANALOG
INPUT menu 2/3.
2) Press F2 or F3 button to select SCALE MIN, then press
[
EDIT –> –>
]
(F6) button to enable editing.
3) Press F2 through F5 button to change the value to
0.000, then press
[
ENTER
]
(F6) button.
4) Press F2 or F3 button to select SCALE MAX, then press
[
EDIT –> –>
]
(F6) button to enable editing.
5) Press F2 through F5 button to change the value to
10.000, then press
[
ENTER
]
(F6) button.
• CHANNEL/ANALOG INPUT menu 3/3
1) Press F4 or F5 button to display CHANNEL/ANALOG
INPUT menu 3/3.
2) PRESENT VALUE displays current measured value.
Use ZERO/SPAN to adjust the measured value.
3) Press
[
RETURN
]
(F1) button twice to return to SET-
TING MENU.
■
PEN SETTING
1) Press F2 or F3 button to select PEN in the menu list,
then press
[
ENTER
]
(F6) button to display PEN SET-
TING menu.
2) Confirm that P1 is displayed at PEN column.
3) Press F2 or F3 button to select INPUT, then press
[
EDIT
–> –>
]
(F6) button to enable editing.
4) Press F2 or F3 button to select “AI 1”, then press
[
EN-
TER
]
(F6) button.
5) Press
[
RETURN
]
(F1) button twice to return to TREND
display screen.
Basic procedure is mostly the same as the above example.