 HOW TO MOUNT THE UNIT ON A PANEL

1) Insert the unit into the panel cutout.  The watertight 

packing must be in place to hold the unit.  Do not remove 

it even when the protection is not required.

  The IP65 front cover protection is effective only with sin

-

gle mounting, with the designated cutout size (31.5 

×

138 

mm).

Panel

Watertight Packing

2)  Push the mounting brackets into the grooves on top and 

bottom of the rear module, until they hit the panel’s rear 
side.

  Leave at least 3 centimeters both above and below the 

unit for ventilation.

EXTERNAL DIMENSIONS

  unit: mm (inch)

75 (2.95)

103 (4.06)

10 (.39)

18

(.71)

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16

16–M3 SCREW

TERMINALS

2–M3 SCREWS

144 (5.67)

137.5 (5.41)

*

1

(.16)

36 (1.42)

31 (1.22)

100 (3.94)

Rounded corners for the option /D

 

OPTION /D BEZEL

*

3

*

1. Space required when replacing the scale plate.

*

2. Used for the existing panel cutout of M-System 48 Series (38 ×

 

139.5 mm).

*

3. Used for the existing DIN panel cutout (33 × 138 mm)

*

4. Used for the existing panel cutout of Fuji Electric PAJ, PAK, PBA (44 ×

 

138 mm), etc.

 

STANDARD BEZEL

*

2

10.4
(.41)

7.5
(.30)

3.9

(.15)

136.5 (5.37)

144 (5.67)

11.4
(.45)

7.5
(.30)

4.9
(.19)

136.5 (5.37)

144 (5.67)

8.5

(.33)

10.4

(.41)

7.5

(.30)

136.5 (5.37)

144 (5.67)

 

OPTION /F BEZEL

*

4

Bracket

Summary of Contents for 48NDR

Page 1: ...invalidate the CE conformance POWER INPUT RATING OPERATIONAL RANGE Locate the power input rating marked on the product and confirm its operational range as indicated below AC rating 85 264V 50 60 Hz approx 5 5 8VA DC rating 24V 15 approx 3 5W ENVIRONMENT Indoor use If the unit s environmental protection is compromised e g when multiple units are to be mounted side by side when heavy dust or metal ...

Page 2: ...ction 138 1 0 L Panel thickness 1 6 8 0 mm L 31 5 36 N 1 N number of units 1 0 CLUSTERED MOUNTING no ingress protection 1 0 USING BEZELS Bezels are used when the unit needs to be adapted to an ex isting panel cutout It is not required when the panel cutout size is as specified in the unit s data sheet 31 5 138 mm Three types of bezels are selectable when ordering Standard type Fits into M System s...

Page 3: ...ze 31 5 138 mm Panel Watertight Packing 2 Push the mounting brackets into the grooves on top and bottom of the rear module until they hit the panel s rear side Leave at least 3 centimeters both above and below the unit for ventilation EXTERNAL DIMENSIONS unit mm inch 75 2 95 103 4 06 10 39 18 71 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 16 M3 SCREW TERMINALS 2 M3 SCREWS 144 5 67 137 5 5 41 4 1 16 36 ...

Page 4: ...xisting panel cutout of Fuji Electric PAJ PAK PBA 44 138 mm etc STANDARD BEZEL 2 10 4 41 7 5 30 3 9 15 136 5 5 37 144 5 67 11 4 45 7 5 30 4 9 19 136 5 5 37 144 5 67 8 5 33 10 4 41 7 5 30 136 5 5 37 144 5 67 OPTION F BEZEL 4 12 13 14 15 16 U V La Lb Lc 12 13 14 15 16 U V LLa 11 La LLb Lc LLc LOAD Inductive Load Coil Varistor or CR Circuit DC Powered LOAD Inductive Load Coil Diode Varistor or CR Cir...

Page 5: ...de 1 orange orange green orange Input L L Input H H Input green red H L Pattern 2 model suffix code 2 orange green green red red red H L 50 40 30 20 10 0 10 20 30 40 50 C C 100 SET H S P Z L 90 80 70 60 50 40 30 20 10 0 SET ENTER Button H Setpoint Adj Trip Indicator red Span Adj amber Decimal Point Position Adj amber Zero Adj amber L Setpoint Adj Trip Indicator green UP Button Mode Selector DOWN B...

Page 6: ... Delay for L Setpoint Increase Decrease Complete Digital Indicator SET L LL M Press at once ON Time Delay for LL Setpoint Increase Decrease Complete Digital Indicator SET LL Z M Press at once Zero Adjustment Digital Indicator Bargraph Z SET Z M Increase Decrease Zero Adjustment Digital Indicator Only Complete Z ON LED OFF SPAN ADJUSTMENT Apply 100 input signal before adjustment All alarm setpoints...

Page 7: ...ue and press SET DIGITAL INDICATOR ZERO ADJUSTMENT 1 Apply 0 input signal 2 Hold down M for 3 seconds and push M to choose Z 3 Press UP or DOWN buttons until the digital indicator comes to a desired value and press SET BARGRAPH DIGITAL INDICATOR SPAN ADJUSTMENT 1 Apply 100 input signal 2 Hold down M for 3 seconds and push M to choose S 3 Press UP and DOWN buttons at once Blinking S LED is switched...

Page 8: ...ing on the digital indicator 3 Press SET and DOWN buttons at once The M button remain activated for 1 minute after it has been used to choose a indicator LED programming mode During this period pressing the M button briefly is enough to choose another indicator After 1 minute it turns to de activated state CALIBRATION PROCEDURE This unit is calibrated at the factory to meet the ordered specificati...