• Normal Scaling
The display value increases when the input signal increases.
• In
ve
rted Scaling
The display value decreases when the input signal increases.
• ‘Teach’ Calibration
‘Input scaling value A’ and ‘Input scaling value B’ can be adjusted by applying actual input signals. These settings will be lost
after an initialization.
How to ‘Teach’ Input Values
1) Hold down Scale/
button for 3 seconds or more to enter the scaling setting mode.
2) Choose an appropriate input type and press Alarm/ button to move on to Input scaling value A.
3) ‘Zro’ and ‘Tch’ indicators turn on.
4) Zero point: Press Up button and ‘Tch’ indicator starts blinking.
5) Apply 5% input signal. Press Up button again and ‘Tch’ indicators returns to ON.
6) Press Alarm/ button to move on to Display scaling value A.
7) Press Alarm/
button again to move on to Input scaling value B.
8) ‘Spn’ and ‘Tch’ indicators turn on.
9) Full-scale point: Press Up button and ‘Tch’ indicator starts blinking.
10
) Apply 100% input signal. Press Up button again and ‘Tch’ indicators returns to ON.
Note 1: For setting input scaling value A and B, proceed with the same method. The combination of inputting parameter and
Teach Calibration is unable.
Note 2: The Teach Calibration is a function to calibrate by applying actual input signal of 5% and 100%. The error might
increase depending on the way of the Teaching Calibration.