background image

EM-9421-B Rev.4

P. 2 / 3

3. INSTALLATION

Keep the ambient temperature and relative humidity within
the limit described in Chapter 2.  10 to 35

°

C (50 to 95

°

F) and 20

to 70% RH are recommended conditions for higher accuracy
requirements.
Insert the meter in pressing the both sides' retainers.

4-2) ANALOG INPUT

Connect DC input to the pin 1 - 2 of CN1.  Typical configuration
is shown in Figure 4.

Figure 4.  Typical connection

4-3) DECIMAL POINT POSITION

Connect terminals as shown below.

DP1 (

×

10

1

) LED ON

Connect CN2 pin 4 and pin 3

DP2 (

×

10

2

) LED ON

Connect CN2 pin 5 and pin 3

DP3 (

×

10

3

) LED ON

Connect CN2 pin 6 and pin 3

4-4) DISPLAY HOLD

When HOLD signal is applied, the display is held.  With contact
input, connect a relay to the pin 1 - 2 of CN2 as shown in Figure
4.  With the contact closed, the display indication is held.  For
other interfaces, examples are shown in Figure 5.

48 (1.89)

45 (1.77)

24 (0.94)

4.5

(0.18)

72 (2.83)

51 (2.01)

21.8 (0.86)

Figure 2.  External dimensions & panel cutout

DP3

1

2

3

4
5

6

7

8

9

D. GND

HOLD

DP COM

DP1

0V

10

24V

DP2

NC

FG

NC

1

2

3

4
5

6

7

8

9

Hi

Lo

NC

NC

NC

10

NC

NC

NC

A. GND

CN2

CN1

10

3

10

2

10

1

DP3

(LED ON)

DP1

10

0

DP2

L

HOLD

twisted-pair

cable

(a) 5V-CMOS, LSTTL

H

(b) TRANSISTOR

L

HOLD

twisted-pair

cable

(c) PHOTO-COUPLER

H

photo-

coupler

D.GND

HOLD

D.GND

HOLD

Figure 5.  Hold control signal interface

45

+0.6 (1.80)
–0 (1.77)

22.2

+0.3 (0.89)

–0 (0.87)

PANEL CUTOUT

Panel thickness:  0.5 to 4 mm
(0.02" to 0.16")

4-1) POWER & FRAME GROUND

Supply DC voltage (rating 

±

10%) to the pin 9 - 10 of CN2.  Use

the pin 8 (FG) of CN2 for grounding when necessary.  This FG
pin is connected to power (0V) through a capacitor (1000PF).

1
2
3
4
5
6
7
8
9

10

NC

DPCOM

D.GND

HOLD

DP1

DP2

DP3

FG

0V

+24V

DECIMAL
POINT

POWER
SOURCE

KEY

*

Pin No.2 (Lo) – 5 (A.GND) of CN1 and No.1 (D.GND) of 

  CN2 are connected internally.
  They are isolated from No.9 (0V) of CN2.

**

The connectors are removable.

CN2

(lower)

1
2
3
4
5
6
7
8
9

10

Hi

Lo

NC

NC

A.GND

NC

NC

NC

NC

NC

INPUT

KEY

CN1

(upper)

ISOLATION

4. TERMINAL CONNECTIONS & ADJ.

Figure 3.  Connector pin assignment