• Place the lamp cap in the original position and rotate it clockwise and insert
the connecting plugs.
• Turn on the main power switch and check whether the new bulb works or not.
4.2 Replacing the reflecting mirror
• Set the angle between the microscope and the illumination arm to exceed 30º
• Remove the long mirror by holding the extended surface.
• Insert new long reflecting mirror.
4.3 Replacing the fuse
• Turn off the main power switch (8) and pull out the input cable from the power
socket.
• Screw off the fuse holder cover with the screw driver (P).
• Replace it with a new fuse, then fasten the cover.
• The fuse specifications are as follows
110V 1A, 250V
220V 0,5A, 250V