46
EMC GUIDANCE
Table 6 – Recommended separation distances between portable and mobile RF communications equipment
and the ME equipment or ME system – for ME equipment and ME systems that are not Llife-supporting.
Recommended separation distances between
portable and mobile RF communications equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the device
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
Not applicable
0.117
0.233
0,1
Not applicable
0.369
0.738
1
Not applicable
1.167
2.333