EMC GUIDANCE
Table 4 – Guidance and manufacturer’s declaration – electromagnetic immunity – for ME equipment and
ME systems that are not life-supporting.
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Not applicable
3 V/m
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of
the transmitter. See recommended
separation distance on next page.