7. OFFSET ADJUSTMENT
7-1 Type K, Type J offset adjustment
1)Set the function to Type K ( Type J ).
2)Insert the probe to the T1 input socket.
3)Pressing " Offset button " ( 3-8, Fig. 1 )
continuously at least two seconds then release,
the display will show :
oFS
SP-t
4)Pressing " L button " ( 3-5, Fig. 1 ) once, the display
will show as example
oFS
23.7
23.7
If not insert probe
to T1 input socket
measuring
adjust
it will show " Err "
value
value
* Use the " button "( 3-5, Fig. 1 ) " button " ( 3-7,
▲
▼
Fig. 1 ) to adjust the desiring value on right bottom
display.
* Pressing " Enter button " ( 3-4, Fig. 1 ) once, both
bottom display will flashing 4 times then return to
normal measuring screen and finish the offset
adjustment procedures.
* The above offset adjustment for Type K ( Type J ) is
valid for T1, T2, T3, T4 at the same time.
16