IMPORTANT:

The compressor motor has a thermal protector which will shut off the unit before 

it overheats. When the thermal protector shuts the unit off:

  1 . 

Switch off the LUMINEB II .

  2 . 

Unplug the unit from the electrical outlet .

  3 . 

Wait 30 minutes for the motor to cool down before another treatment .

  4 . 

Make sure the air openings are not obstructed .

CLEANING YOUR LUMINEB II

  It is recommended that the nebulizer, mouthpiece and mask are thoroughly 

cleaned with hot water after each use and cleaned with mild detergent after the 
last treatment of the day . If your physician or respiratory therapist specifies a 
different cleaning procedure, follow their instructions .

  Before cleaning, make sure the power cord is disconnected from the electrical 

outlet and the unit .

  To clean your LUMINEB II:

  1 . 

Disconnect the air tube, nebulizer and mouthpiece .

  2 . 

Gently twist the nebulizer to open it .

  3 . 

Rinse the nebulizer and mouthpiece or mask with water .

  4 . 

Dry with a clean soft towel or allow to air dry .

5 .    Reassemble the nebulizer when completely dry . 
6 .    Wipe the compressor with a damp cloth and allow to air dry .

  7. 

Never use powdered cleaners or soap pads on the compressor, as it may

 

 

damage the finish of the unit .

DISINFECTING YOUR LUMINEB II

It is suggested that your LUMINEB II be disinfected after the last treatment of 

each day. Unless indicated otherwise by your physician, disinfect your LUMINEB II:

Summary of Contents for LUMINEB II 5710

Page 1: ...Model 5710 LUMINEB II Compressor Nebulizer Operation Instructions 5710 INS LAB RevB14...

Page 2: ...SPECIFICATIONS 10 FIVE YEAR LIMITED WARRANTY 11 INSTRUCCIONES EN ESPA OL 13 GF Health Products Inc is not responsible for typographical errors For the most updated and current information on packagin...

Page 3: ...ry disorders It creates a stream of air that travels through clear tubing to the nebulizer When air enters the nebulizer it converts the prescribed medication into aerosol mist for easy inhalation You...

Page 4: ...ces Disconnect the unit from the electrical source before cleaning filling the nebulizer and after each use Ensure the power cord is fully inserted into the socket to prevent electrical malfunction Cl...

Page 5: ...Nebulizer Compressor 4 Tube Connector 7 Power Cord 10 Power Cord Plug 2 Air Tube 5 Standard Mouthpiece 8 Air Vents 11 Filters 3 Filter Compartment 6 Nebulizer 9 On Off Switch PARTS IDENTIFICATION OPTI...

Page 6: ...seated 2 Gently twist the top part of the nebulizer assembly counter clockwise to disassemble the nebulizer fig 1 3 Fill the bottom section of the nebulizer with the medication prescribed by your phy...

Page 7: ...the top of nebulizer assembly fig 5 If using an optional mask attach the mask to the top of nebulizer assembly fig 6 8 Make sure the On Off switch is in the Off position then plug the power cord into...

Page 8: ...ry therapist specifies a different cleaning procedure follow their instructions Before cleaning make sure the power cord is disconnected from the electrical outlet and the unit To clean your LUMINEB I...

Page 9: ...tion for thirty minutes CHANGING THE FILTER 1 Use only Lumiscope nebulizer filters 2 Do not substitute cotton or any other materials Do not wash clean or re use a filter 3 Never operate your LUMINEB I...

Page 10: ...umption 70 Watts Medication Capacity 5 ml Particle Size 4 m Sound Level 60dba Maximum Medication Capacity 13ml cc Average Nebulization Rate 0 2ml min Compressor Pressure Range 30 36 Psi 210 250 KPa 2...

Page 11: ...handling Please make checks payable to GF Health Products Inc Failure to follow the specificdirections provided by the GF customer service team will result in denial of thewarranty claim ENTIRE WARRAN...