4.8 Performing a Quality Control test
This Platform User Manual describes the basic Quality Control solution test method for
the LumiraDx platform. If there are differences in the test method for a particular QC test,
the Instrument will prompt at each step of the process. Please refer to the LumiraDx Test
Strip Product Insert before performing a new QC test for specific information.
Important Notes
ALWAYS
•
Wear a clean pair of gloves before performing a QC test.
•
Place the Instrument on a level, stable surface before starting a test.
•
Use only LumiraDx Quality Control solutions.
•
Apply Quality Control solution
AFTER
inserting the Test Strip and the
Instrument prompt.
•
Follow the test operating conditions in the LumiraDx Test Strip Product
Insert.
•
Follow the information on correct handling of Test Strips in the LumiraDx
Test Strip Product Insert.
Instrument Operation - 45
To view test information during a QC test, or on the test result screen, tap the at the
top of the touch-screen.
To cancel a QC test tap ‘Cancel’ at the top of the touch-screen and follow the
Instrument prompts.