Lumina Power XLB-3000 User Manual Download Page 7

 

 

 

XLB-3000-XX-YY Specifications 

XX = Iout

max

   YY= Vout

max   

  

 

Model 

P

out max

 

I

out max

 

Input Voltage 

Size (L x W x H) 

XLB-3000-XX-YY 

3000W 

Configurable to 

120A 

 

200-240VAC 

13” x 8.55” x 3.43” 

33cm x 21.7cm x 8.71cm 

Max Lamp Voltage:  

35V                                    

Auxiliary Outputs

:   +15V @ 0.20A

 

 

Input 

Voltage:  

 

200-240VAC, 50/60 Hz, 18.5A @ 208VAC 

Power Factor: 

 

>.98 

 

Interface

    

(See interface description page 3) 
Connector: 15 Pin D-Type Sub Female

 

 

Ignition/Boost 

Boost Voltage:   

250V 

Boost Energy:    

500 mJ 

Ignition Voltage:  

Up to 45kV (~1us rise time) 

Igniter Polarity:  

Positive or Negative (Factory Set per customer request) 

Ignition Energy:  

65mJ 

Igniter Dimensions:  

5.5” x 3.6” x 2.6” 
140 x 92 x 66mm 

Performance 

Line Regulation:   

<0.2% of maximum output current 

Current Regulation:  

<0.5% of maximum output current 

Current Ripple:    

<0.5% of maximum output current 

Power Limit:  

 

Limited to maximum power with powerfold-back circuit 

 

Environment 

Operating Temp:  

0 to 40

o

Storage:  

 

-20 to 85

o

Humidity: 

 

0 to 90% non-condensing 

Cooling:  

 

Forced air 

 

Regulatory 

Leakage Current:  

<350uA 

 

Approvals:   

 

 

Medical Safety:   

UL60601-1, IEC 60601-1, EN 60601-1, CAN/CSA C22.2 No. 601.1-M90 

 

Emissions/Immunity: 

FCC 47 CFR Class A Emissions, EN55011:1998 Group 1 Class A Emissions, 
EN61000-3-2 Limits for harmonic current emissions, EN 610000303 Flicker, 
EN60601-1-2:2001 Electromagnetic emissions and immunity for medical 
equipment 

 
 
 
 

Summary of Contents for XLB-3000

Page 1: ...1 USER MANUAL XLB 3000 Short Arc Xenon Lamp Ballast...

Page 2: ...les and Figures Figure 1 XLB 3000 Block Diagram 6 Figure 2 XLB 3000 Chassis Outline Drawing 8 Figure 3 XLB 3000 Igniter Drawing 9 Table 1 XLB 3000 Interface 10 Figure 4 XLB 3000 Interface Schematic 11...

Page 3: ...tion and industrial applications where a stable light source is essential As a Xenon lamp ballast the XLB 3000 power supply first ignites the lamp with a high voltage pulse and once the lamp is ignite...

Page 4: ...e personnel of Lumina Power Inc are qualified to service this equipment High Voltage Present This power supply produces lethal high voltages Only service personnel of Lumina Power Inc are qualified to...

Page 5: ...aving a high harmonic content With power factor correction the non 50 60 Hz harmonics are reduced to near zero Since only the fundamental frequency is now used to deliver power the efficiency of the p...

Page 6: ...ll internal power supply requirements as well as the external 15V and 5V power supplies are derived from the power factor control boost inductor All auxiliary power supplies are regulated by standard...

Page 7: ...st Ignition Energy 65mJ Igniter Dimensions 5 5 x 3 6 x 2 6 140 x 92 x 66mm Performance Line Regulation 0 2 of maximum output current Current Regulation 0 5 of maximum output current Current Ripple 0 5...

Page 8: ...8 Figure 2 XLB 3000 Chassis Outline Drawing For 3 D model consult factory...

Page 9: ...9...

Page 10: ...t of the supply can be monitored by Iout Monitor 7 Iprog input 0 10V 0 Iout max The power supply output current is set by applying a 0 10V analog signal toIprog Note that with I prog set between 0V an...

Page 11: ...11 XLB 3000 XX YY Interface Schematic Figure 4...

Page 12: ...the leads between the igniter and the Xenon lamp as short as possible No other wires should be in the vicinity of the igniter output wires connecting to the Xenon lamp The HV pulse produces transient...

Page 13: ...13 NOTE Negative ignition is also available Figure 6 XLB 3000 Lamp Connections...

Page 14: ...before applying AC Power The unit may be programmed for output current via Pin 7 the I prog function But there are three interface control signals which must be properly set before the output will de...

Page 15: ...Interlock Pin 3 is connected to GND via Interface pins 4 9 or 15 c Apply Input AC Power Turn ON AC power After a few seconds the power supply fans will begin to run d Programming Output Current Progra...

Page 16: ...the XLB 3000 XLB 3000 units have no serviceable parts Do not attempt to repair or service this unit in the field Removing tamper seals from chassis will void warranty For further information contact L...

Reviews: