EN
34
Test voltage:
2210 V a.c. rms 1 minute between housing / power supply and:
- RS485; - binary outputs; - USB; - analog inputs
1390 V a.c. rms 1 minute between:
- analog inputs / RS485; - analog inputs / binary outputs
- analog inputs / USB; - USB / RS485
- USB / binary outputs; - RS485 / binary outputs
Degree of protection IP:
from the front
IP 50
from the terminals IP 00
Power consumption:≤ 3 VA
Weight
< 0.2 kg
Dimensions
53 X 110 X 60 mm
Nominal operating conditions:
- supply voltage
85...253 V a.c. 40..400 Hz; 90...300 V d.c.
20...40 V a.c. 40...400 Hz, 20...60 V d.c.
- ambient temperature
-10...23...+55 °C
- storing temperature
-25...+70 °C
- humidity
< 95% (without condensation)
- external magnetic field
0..40..400 A/m
- operation position
vertical
- warm-up time
30 min.
EMC compatibility:
- immunity to interference in accordance with EN 61000-6-2
- interference emission in accordance with EN 61000-6-4
Safety requirements:
in accordance with the standard EN 61010-1
•
insulation between circuits basic
• installation category
III,
• degree of pollution
2,
• maximum voltage relative to earth:
- for power circuit 300 V
- for other circuits 50 V
• altitude < 2000 m