EN
32
4. TECHNICAL DATA
Input ranges:
Input voltage:
voltage (maximum range): -12...-10...0...10...12 V
internal resistance: > 1 MΩ
basic error: 0.1 % of the range
Current input:
current (maximum range): -24...-20...0...20...24 mA
internal resistance: 10 Ω ±1%
basic error: 0.1 % of the range
Pt100 sensor input:
range of measure temperature: -200 °C…850 °C
basic error: 0.1 % of the range
sensor current: 300 µA ±2%
maximum lead resistance: < 20 Ω
Pt500 sensor input:
range of measure temperature: -200 °C…850 °C
basic error: 0.1 % of the range
sensor current: 300 µA ±2%
maximum lead resistance: < 20 Ω
Pt1000 sensor input:
range of measure temperature: -200 °C…850 °C
basic error: 0.1 % of the range
sensor current: 300 µA ±2%
maximum lead resistance: < 20 Ω
Voltage input (shunt):
voltage (maximum range): -180...-150...0...150...180 mV
internal resistance: > 50 kΩ
basic error: 0.1 % of the range
J type thermocouple input:
range of measure temperature: -210 °C…1200 °C
basic error: 0.1 % of the range
K type thermocouple input:
range of measure temperature: -250 °C…1372 °C
basic error: 0.1 % of the range