Transport, packaging, storage 

 
The instrument can be damaged or destroyed if shipped incorrectly. To transport or store the 
instrument, please use the original box or a box padded with sufficient shock-absorbing material. 
For storage in humid areas or shipment overseas, the device should be placed in welded foil 
(ideally along with silica gel) to protect it from humidity. 
The pyrometer is designed for a storage temperature of -20 … 70°C with non-condensing condi-
tions. Other kind of storage can damage or malfunction the pyrometer.  
 
 

8

 

Maintenance 

 

8.1 

Safety 

Attention during pyrometer services:  
Should the pyrometer be integrated in a running machine process the machine has to be switched 
off and secured against restart before servicing the pyrometer. 
 
 

8.2 

Service 

The pyrometers does not have any parts which require regular service, only the lens has to be 
kept clean. The lens can be cleaned with a soft cloth in combination with alcohol (do not use acid 
solutions or dilution). Also standard cloths for cleaning glasses or photo objectives can be used. 
The Germanium lens of the IN 300 has an anti-reflective coating which appears slightly colored. 

 

Be extremely careful - this layer can easily be rubbed off - this will greatly affect the measuring 
results! 
 
 

Summary of Contents for IN 300

Page 1: ...IN 300 IN 300 5 IMPAC Pyrometer Betriebsanleitung Operation Manual Proven Quality www imaxsun com...

Page 2: ...ons 15 2 2 Appropriate use 16 2 3 Scope of delivery 16 3 Safety 16 3 1 General 16 3 2 Electrical connection 16 4 Electrical Installation 16 4 1 Accessories 17 5 Optics 18 5 1 IN 300 18 5 2 IN 300 5 18...

Page 3: ...nd experience LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be incom...

Page 4: ...tput 4 20 mA load independent current linear to temperature Max load max 500 at 24 V max 200 at 18 V Emissivity 0 4 1 adjustable Repeatability 1 of the measuring range 1 Tamb 23 C Accuracy 1 5 of temp...

Page 5: ...connection Follow common safety regulations for mains voltage 230 or 115 V AC connecting additional devices operating with this mains voltage e g transformers Touching mains voltage can be mortal A no...

Page 6: ...C Recorder Digital display 230 V 4 20 mA white brown Shield black 24 V DC Power supply Wiring white 24 V brown 0 V black Shield Note Additional analyzing instruments e g controllers recorders etc can...

Page 7: ...the measuring distance in mm e g 300 mm in which it has the smallest spot size in relation to its measuring distance IN 300 e g 20 mm The following drawings show the size of the spots in mm in depende...

Page 8: ...is set ex works to an emissivity of 100 To adjust the emissivity factor to your own purpose you have to remove the rear cover by unscrew the locking screw Remove the cover carefully In the tube there...

Page 9: ...unction the pyrometer 8 Maintenance 8 1 Safety Attention during pyrometer services Should the pyrometer be integrated in a running machine process the machine has to be switched off and secured agains...

Page 10: ...8 2 Temperature indication too high Emissivity set too low Set lower correct emissivity corresponding to the material see 6 The measurement is influenced by reflections of hot machine parts Use mecha...

Page 11: ...DA 4000 LED display 2 wire power supply 2 limit switches relay contacts 3 890 520 DA 6000 LED display RS232 2 wire power supply maximum value storage analog output 3 890 530 DA 6000 LED display RS485...