Series IN 500 Operating Manual
Introduction 10
Digital interface:
Switchable: RS232 or addressable RS485 (half duplex)
baud rate 1200 up to 19200 Bd, resolution 0.1 °C
Isolation:
Non isolated analog outputs and power supply (relay contact isolated)
Emissivity (EMI):
10 to 120% adjustable in steps of 0.1%
Max. / Min. value
storage:
Clear time OFF; 0.1 s; 0.25 s; 0.5 s; 1 s; 5 s; 25 s; extern; auto
Response time t
90
:
180 ms; switchable: 0.5; 1; 2; 5; 10 or 30 s
Temperature display
(only IN 510 and IN 520):
LCD, 4 digit, 3 values per second, permanent display illumination
Exceeding of measuring
range:
Display: 8888
Fall below measuring
range:
Display: 1 °C below sub range
Resolution:
1/10 °C (1/10 °F from -40 to 999.9 °F; 1 °F from 1000 to 1292 °F)
Uncertainty
**)
Dependent on
object temperature T
and ambient
temperature T
amb.
(EMI = 1, t
90
= 1 s):
**)
With thermocouple
output minimum 2.5 °C
T = 0 to 700 °C: 0.8% or 1 °C
*)
T = 0 to -20 °C: 2 °C
T = -20 to -40°C: 3°C
T
K
of uncertainty: 0.03%/°C or 0.05°C/°C at 25 °C ambient temperature
T
amb.
*)
*)
Whichever value is greater. The instrument must be at a constant
ambient temperature for a minimum of 15 minutes.
A correct temperature measurement is impossible if the
temperature of the object is more than 85 °C below the sensor head
temperature (temperature reading is too high, no error message)
Repeatability:
0.5% of measured value °C or 0.5 °C
whichever value is greater, ambient temperature is constant
Noise Equivalent
Temperature Difference
(NETD):
with t
90
= 180 ms: 0.1 °C ( = 1)
(measured temperature = 23 °C and emissivity = 1)
Ambient temperature
converter:
0 to 65 °C
Storage temperature
converter:
-20 to 70 °C
Ambient temperature of
sensor head:
IN 510-N and IN 510: 0 to 85 °C
IN 520-N and IN 520: 0 to 180 °C
with cooling / purging unit:
0 to 200 °C
Storage temperature of
sensor head:
-20 to 85 °C
(IN 510-N, IN 510)
-20 to 180 °C (IN 520-N, IN 520)