Series IN 500 Operating Manual
Introduction 9
2 Introduction
2.1 Appropriate use
The
IN 510-N, IN 510, IN 520-N, and IN 520
are stationary pyrometers for non-contact
temperature measurement of non-metallic surfaces or painted, coated or anodized metals.
The versions IN 510 and IN 520 are equipped with an illuminated LC display which shows the
actual temperature reading. All available parameters can be set via the integrated keyboard.
The types IN 510-N and IN 520-N do not have display and keyboard, they will be parametrized
via the interface.
Type 510 is equipped with a sensor head and sensor cable for an ambient temperature of max.
85 °C without cooling, type 520 is equipped with a sensor head and sensor cable up to
180 °C without cooling.
2.2 Scope of delivery
Instrument, one sensor head with cable length of 3 m or 15 m cable, works certificate, and
operation manual.
Note:
A connection cable is not included with the instrument and has to be ordered
separately (see section
8, Reference numbers
).
2.3 Technical Data
Temperature range:
-40 ... 700 °C (-40 ... 1292 °F)
Sub range:
Factory setting: 0 ... 500 °C; user adjustable (minimum span: 51 °C)
Data handling:
Digital
Spectral range:
8 to 14 µm (for measurement of non-metallic or coated metallic
objects)
Sensor head:
Optics 10:1: with lens; optics 2:1: without lens
IR detector:
Thermopile
Power supply:
10 to 30 V DC ripple < 0.5 V
power consumption: max. 60 mA
Analog output:
Linear current (0/4 to 20 mA), voltage (0 to 5 V) or
thermocouple (type “J“ or “K“)
Additional output:
10 mV/°C or 10 mV/°F for temperature of sensor head
Load:
Max. 700 / 24 V power supply (for current output) (500 /20 V)
Output impedance:
100 (for thermocouple or voltage output)
Relay contact:
Isolated relay contact; 50 V DC; 0.2 A
Hysteresis:
Negative hysteresis, 2 – 20 °C adjustable
(without current or value exceeded = open contact)