After making the gain Calibration please make sure that the Mode is set
back into 5 and that the Zero calibration is set to “Auto”, as described in
4.3.8 Installing a new NO sensor.
After installation the new NO sensor it must be span calibrated following
the procedure described in the previous sections.
After performing the NO span calibration:
Select the New NO sensor in the Bench command menu, see figure
4.35.
Figure 4.35. Bench command menu select New NO sensor
The following window appears.