Model 2360
Technical Manual
Section 4
Ludlum Measurements, Inc.
Page 4-4
July 2016
SWITCH
DIVIDE BY
3 4
RATIO
ON ON
1
ON OFF
10
OFF ON
100
OFF OFF
Beta audio
Subtract Mode
LOG SAMPLE
: A two-pole DIP switch (#5 & #6) used to select logging
mode. When both switches are
OFF
, the instrument will not log samples
but will output the ratemeter readings to the RS-232 port every two
seconds in ASCII format. When set to log both scaler and ratemeter, two
samples are actually saved. The logging mode is selected from the
following table.
SWITCH
LOGGING
5 6
MODE
ON ON
Log ratemeter and
scaler
ON OFF
Log
ratemeter
OFF ON
Log
scaler
OFF OFF
No sample saved.
Output ratemeter every
2
seconds.
The following controls are utilized during calibration only and should
only be altered by a qualified calibration technician.
MTR
: a multi-turn potentiometer used to calibrate the meter to the
CPM
reading
AT
: a multi-turn potentiometer used to vary the alpha pulse discriminator
from 40 to 700 millivolts (mV)
BW
: a multi-turn potentiometer used to vary the upper beta pulse
discriminator from
BT
setting to
AT
setting
BT
: a multi-turn potentiometer used to vary the lower beta pulse
discriminator from 2 to 15 mV
OL
: a multi-turn potentiometer thatprovides a means to vary the detector
current overload set point
Summary of Contents for 2360
Page 4: ......
Page 40: ...Model2360 TechnicalManual Section7 Ludlum Measurements Inc Page 7 14 July 2016 ...
Page 41: ...Model2360 TechnicalManual Section7 Ludlum Measurements Inc Page 7 15 July 2016 ...
Page 42: ...Model2360 TechnicalManual Section7 Ludlum Measurements Inc Page 7 16 July 2016 ...
Page 59: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......