
40
CHAPTER 4. SETUP MODE
• Ones Place (0-9)
• Tens Place (0-9)
• Hundreds Place (0-9)
Beta RATE/MAX Mode Alarm Point (Default 5.00 kcpm) -
A short press of ON/OFF to select
the value to adjust, and MODE to adjust the value. Units will be the same as selected earlier
with RATE/MAX Units (P2-1). The ALARM LCD segment will be on to indicate an alarm
parameter. Set this Alarm Point to 000 to disable. Available values are:
• Ones Place (0-9)
• Tens Place (0-9)
• Hundreds Place (0-9)
• Number of Decimal Places (0, 1, or 2)
• Range (k on/off )
ð
If the RATE/MAX Units has changed to a value other than that used to previously
set this Alarm Point, the Alarm Point will be reset to 000.
Beta Count Alarm Point (Default 5.00 kc) -
A short press of ON/OFF to select the value to
adjust, and MODE to adjust the value.Units will be the same as selected earlier with Count
Units (P2-3). The ALARM LCD segment will be on to indicate an alarm parameter. Set this
Alarm Point to 000 to disable. Available vallues are:
• Ones Place (0-9)
• Tens Place (0-9)
• Hundreds Place (0-9)
• Number of Decimal Places (0, 1, or 2)
• Range (k on/off )
ð
If the COUNT Units has changed to a value other than that used to previously
set this Alarm Point, the Alarm Point will be reset to 000.
Beta Loss of Count Alarm Time (Default 60 seconds) -
A short press of ON/OFF to select
the value to adjust, and MODE to adjust the value. Setting the Loss of Count Alarm Time to
0 disables the alarm. Value is in seconds. Available values are:
Ludlum Measurements, Inc.