Model 2241-2RK
Technical Manual
Section 5
Ludlum Measurements, Inc.
Page 5-4
July 2012
D
ETECTOR
M
ODEL
44-9
T
UBE
R
EPLACEMENT
P
ROCEDURE
Consult the Model 44-9 Alpha-Beta-Gamma Detector drawing in Section 11
of this manual to perform the following:
Caution!
The mica window of this tube is
extremely thin
and can easily
break. There is also a thin layer of material to prevent UV
interference. This material may come off if touched, causing
the detector to malfunction. DO NOT TOUCH.
1.
Remove the back plate by removing the three screws.
2.
Loosen the three set screws on the side of the tube housing.
3.
Remove the old tube from the detector housing.
4.
Remove the anode clip from the old tube.
5.
Push the clip onto the anode housing.
Note:
Do not over-flex the wire when installing the clip!
6.
Carefully install the tube with the window face-down in the
housing. Tighten the set screws evenly. The tube should be
flush against the screen.
7.
Replace the back plate and the retaining screws.
8.
The detector is now ready for use.
Note:
The instrument and detector will need to be recalibrated once
the above procedure is complete.
Summary of Contents for 2241-2
Page 3: ......
Page 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Page 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Page 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......