Compare the first reading with the second to determine the
amount of alpha and beta contribution. (This will be a very
rough number.)
Gamma and X-ray
: For gamma and high energy X-ray, the gamma
scintillator (Model 44-2) must be used.
This detector is approximately 50 times more sensitive than
the pancake (Model 44-9) detector. This detector is able to
detect low-level gamma and X-rays in the range of 60 keV-
1.25 MeV. This detector may be used for measurements up
to 50 mR/hr.
The Model 44-2 is used by placing the detector in hand, and
by slowly moving it from side to side over the area to be
measured.
Be aware that the very end of the detector, opposite the
connector, is the most sensitive. Point the detector towards
the radiation being measured.
Summary of Contents for 2241-2
Page 3: ......
Page 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Page 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Page 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......