R-Log – Quick Start 

 

Cod. INSTUM_00953 

Pag. 12 

 

 

 
Besides, if a measure is used by a derived quantity, the icon changes: 

.  

 
The measures order can be changed according to your requirements by pressing the 

Sort

 button. It is 

however  advisable  to  keep  coupled  the  quantities  that  need  to  be  acquired  together  (for  ex.:  wind 
speed and direction) and give priority to the measures with a fast acquisition rate, moving them on 
top of the list.  
 
 

SETTING THE ELABORATION RATE

 

 
The  elaboration  rate  is  10  minutes  by  default.  If  you  wish  to  change  this  parameter,  select 

Elaborations

 from the 

General Parameters 

section. 

 
 

SETTING THE ACTUATION LOGIC 

 

 
The instrument has 2 actuators that can be used for the power supply of the sensors connected to the 
terminal  board.  Actuators  can  be  also  used  by  programmable  actuation  logics,  able  to  generate 
alarms  in  relation  to  the  values  acquired  by  sensors.  The  voltage  available  on  these  terminals 
depends  on  the  power  supply  provided  by  the  instrument.  The  association  between  input  and 
actuator is fixed and follows the table shown in §3.4. 
 
To set an actuation logic, proceed as follows: 
 
 

 

Select 

Logics

 from the 

Actuators

 section; 

 

Select the first available position (for example (1)) and press 

New

 

Select the type of logic from the 

Value 

column, set the requested parameters and press 

OK

 

Select 

Actuators

 from the 

Actuators

 section; 

 

Select the actuator number for association with the logic (for example (7)) and press the 

New 

key; 

 

Enter a check mark in correspondence to the previously entered logic and press 

OK

 

 

SETTING THE OPERATING CHARACTERISTICS 

 
The most significant operating characteristic is the possibility to turn off your display after about one 
minute  of  non-use  so  as  to  reduce  energy  consumption.  It  is  recommended  to  enable  this  option 
when the instrument operates with a battery, with or without PV panels.  
 
Proceed as follows to access the operating characteristics and - in particular - to set the display auto 
shut-off function:  
 

 

Select 

Characteristics

 from the 

Instrument Information 

section; 

 

Select 

Display auto power off

 and set 

Value

 to 

Yes

 
 

SETTING RADIO COMMUNICATION PARAMETERS  

 

Summary of Contents for ELR510M

Page 1: ...Cod MW6165 R Log Quick Start Update 01 25 2016...

Page 2: ...g a configuration report 14 3 4 Connecting the probes 14 3 4 1 Electrical connection 14 3 5 Displaying measures in fast acquisition mode 16 3 6 Starting the system and displaying measurements 16 4 Ela...

Page 3: ...DOCUMENTS section 2 To start There are the following models of R Log Model Connector inputs Terminal board inputs Master mode Repeater mode Slave mode Ethernet output ELR510M X X ELR510R X X ELR510S X...

Page 4: ...R Log Quick Start Cod INSTUM_00953 Pag 4 Legenda ________ cable connection _ _ _ _ _ _ radio connection Fig 1 A type installation Fig 2 B type installation Fig 3 C type installation...

Page 5: ...each Slave Repeater datalogger with 3DOM software Configuration of the Master datalogger with measurements imported by Slave Repeater devices Generation of a Configuration Report for each datalogger C...

Page 6: ...Information Installation in the SOFTWARE section Follow the instructions indicated under Guided Installation of LSI LASTEM programs paragraph 3 1 2 3DOM Software Through 3DOM software you can perform...

Page 7: ...he instrument Insert the instrument in 3DOM Check the instrument s internal clock Create the configuration in 3DOM Send the configuration settings to the instrument For ELR510M ELR510S and ELR510R mod...

Page 8: ...the screen for about 10 seconds when the window disappears the survey will be automatically started based on the last configuration stored before the instrument shutdown Refer to 3 2 4 if the system...

Page 9: ...is mandatory and cannot be cancelled In order to have accurate time data the datalogger internal clock should be correct Failing this the clock can be synchronized with that of your computer through...

Page 10: ...program list Select your instrument serial number in the Instruments panel Select the name of the basic configuration in the Configurations panel user000 by default Press the selected name with the ri...

Page 11: ...sors registry or you want to connect it to the datalogger in single ended mode in this case refer to the instrument user manual press the New button to add a measure entering all the parameters reques...

Page 12: ...the instrument The association between input and actuator is fixed and follows the table shown in 3 4 To set an actuation logic proceed as follows Select Logics from the Actuators section Select the...

Page 13: ...tted via radio to the Master instrument must be entered in the latter as serial measurements To do so proceed as follows Select Measures in the General Parameters section Press Import button to start...

Page 14: ...later when connecting the probes to the datalogger 3 4 Connecting the probes It is recommended to connect the probes with the instrument turned off 3 4 1 Electrical connection For ELR516M and ELR515...

Page 15: ...Rx A B C D Current signal from 4 wires sensor powered fixed or switched from E Log 0 V 12V fixed or switched Probe power supply Rx A B C D Rx Current signal from 2 wires probe powered from signal wir...

Page 16: ...activate the fast acquisition mode proceed as follows Turn the instrument on by simultaneously pressing and buttons Press simultaneously and buttons in the instant values display window Check when pos...

Page 17: ...me based on the selected settings Basic folder Serial number Prefix _ Serial number _ yyyyMMdd_HHmmss txt Note If the setting Append data on the same file is not selected each time instrument data are...

Page 18: ...Pag 18 To enable storage select Check to activate data storage control box The list shows the current connection status This can be changed by pressing the Select key that opens the configuration wind...

Page 19: ...configuration of the programs that use the removed data source and that will continue to use it The timeout for data requests from the database can also be changed To add a new connection select the...

Page 20: ...ted instrument the Download button is enabled proceed then as follows Select the date from which to start downloading data in case some data were already been downloaded the control proposes the date...

Page 21: ...Select the identified acquisition with the starting date time of measurements Press the selected acquisition with the right key of your mouse and choose Show Data for the wind direction measurement s...

Page 22: ...R Log Quick Start Cod INSTUM_00953 Pag 22 To display the chart select Show Chart on the table with the right key of your mouse Enjoy your R Log...

Reviews: