4 Function check
4.2 Checking prismaTSlab
1. Connect up a therapy device and all components for remote
setting (see prismaTSlab instructions for use).
2. Start prismaTSlab.
Requirement
: The therapy device is shown in the list of
detected devices.
3. Select a therapy device from the list of detected devices.
or
In case of 1:1 connection: Check whether the therapy device
connected is selected automatically (see prismaTSlab
instructions for use).
4. Assign a patient to the therapy device.
There is a functioning connection between prismaTSlab and
the therapy device if
• the connected therapy device model and the correct serial
number are displayed in the prismaTSlab information bar.
• you click the
on/off
button
in prismaTSlab and can
start and end the therapy with the therapy device.
• values are displayed when therapy is in progress and you
can change the therapy settings and transfer them to the
therapy device.
If one or more requirements are not satisfied or prismaTSlab is not
functioning correctly, please check that the therapy device and PC
are connected properly
. Restart the device and
prismaTSlab.