10
→
Home care water chamber
When using the Home care water chamber, fill clean water or medical
sterile water (e.g. ALLEGRA sterile water) into the water chamber. By
heating the water to 100°C before use, you reduce limescale deposits
in the water chamber. Do not fill the water chamber above the mark for
the maximum water level.
Note to the reader:
The water chamber is heated during operation.
Please avoid touching the metal base of the water chamber and the
heating plate of the high-flow device.
→
Clinic water chamber (Auto-Fill)
When using the Clinic AutoFill water chamber, hang a sterile water
bag in a high position. Insert the inlet needle into the rubber stopper of
the water bag and open the ventilation opening of the bag. The Clinic
AutoFill water chamber is automatically filled with water to the intended
level.
Please use sterilised water suitable for medical purposes and not more
than 2000 ml.
After connecting the water bag, check that the water flows into the
Clinic AutoFill water chamber and is below the marking line. Replace
the clinic AutoFill water chamber in the event of problems.
Make sure that water does not leak from the Clinic AutoFill water
chamber and water bag during use. Otherwise the Clinic AutoFill water
chamber will run dry and impair the humidification effect.
To avoid damage to the Clinic AutoFill water chamber, avoid running
empty.
→
Therapy tube
Check the therapy tube and nasal cannula before connecting them and
replace them if a breakage or other damage is visible.
Do not expose uncovered skin to the heated therapy tube for a long
time to avoid skin irritation caused by the heat.
Do not run the therapy tube near a heat source (e.g. heater). Do not
cover large areas of the therapy tube. Otherwise the temperature may
rise.
Notes to the reader / Warnings