When going to bed the patient needs to:
•
Attach the MiniScreen Pro using the elasticated
thorax belt.
•
Attach the flow prong and finger sensor.
•
If applicable, attach the abdominal belt, leg
sensor, ECG and EEG electrodes.
•
Check that the sensors and their connectors are
positioned correctly.
•
If manual recording start was selected: Switch
on the MiniScreen Pro by sliding the switch to
the right.
The next morning
:
•
Stop the measurement by switching off the Mini-
Screen Pro.
•
Detach the MiniScreen Pro and sensors and
return all parts to the case. Cleaning of the
sensors will be carried out by the technical staff
at the clinic or hospital.
•
Bring the case back to the clinic or hospital.
•
For automatic evaluation the measurement can
be transferred to the PC using the device
software. Select the "Device connection" tab
followed by "Read measurement from device"
(or select the following tab button
).
3.4. Online Measurements (Inpatient)
Please adhere to the following steps when perfor-
ming an online measurement:
1. Ensure the battery is fully charged (see Page 8).
2. Apply the device and connect all sensors to the
patient (see Page 6)
3. Connect the MiniScreen unit (switched off) to
your PC using the USB interface cable.
4. Start the MiniScreen software on your PC.
5. Complete the Patient data fields in the menu
item ”Device Connection / Online / Record“ and
start the transfer.
6. Switch on the MiniScreen Pro unit by sliding the
switch to the right position. The device can be
switched on also direct after attaching the
device to the patient.
7. Perform impedance test and bio signal calibra-
tion (see the online help of the software).
8. Once the test is complete, (the following mor-
ning) click the stop button
to end the mea-
surement.
9. Switch the MiniScreen Pro device off by sliding
the switch to the left position.
- 7 -