middle of the channel.

Check the flow prong on the patient and the connec-
tion on the MiniScreen unit. When measuring during
CPAP therapy, check the connection of the CPAP
adapter hose on the CPAP mask and on the
MiniScreen Pro unit.

Signal amplitudes are very small or do not
appear.

Check the corresponding sensors on the patient and
their connections to the MiniScreen Pro unit. Whilst
doing so, check that the tube hoses and the black
pressure pads of the thorax sensors are intact. A
leaking sensor can cause inaccurate readings in the
thorax channel. The pressure pads should be
attached

to

the

patient

securely without

over

tightening the belt and the tube hoses should be
routed in such a way that they cannot kink.

Check all cables and connectors on the MiniScreen
Pro unit and the PC.

Pulse oximeters not responding.

The channels for oxygen saturation and pulse
frequency are registering 50% and 30 P/min respec-
tively despite the patient being connected.

First check that the finger sensor is seated correctly
on the patient and remove any nail polish that may
be present. A small red light should be present in the
sensor when the finger is inserted. If the light doesn't
switch on when the sensor is applied to the finger,
then you need to check the connection at the
MiniScreen Pro unit and any intermediate extensions.

Printout not working.

The printer prints characters on the page but in no
apparent format.

The wrong printer or printer driver has been installed.

The printer is not responding to the print command.

Check the printer cable and the connections to the
printer and PC. The printer should be switched on
and operational, i.e. the control LED’s on the printer
should be lit.

Cannot establish connection to the MiniScreen.

The USB cable is not connected correctly.

Check the connection of the USB cable to the
MiniScreen Pro and the PC.

USB interface was deactivated

The

USB

interface

can

be

activated

in

the

MiniScreen

Pro

software.

Select

the

"Device

connection" tab followed by menu "Settings"

.

Battery is discharged.

Charge the battery correctly.

For clarification and in case of problems during
the installation, maintenance or use, please refer
to your local distribution partner, or directly to
the manufacturer.

- 9 -

Summary of Contents for 200-0902

Page 1: ......

Page 2: ......

Page 3: ...ax and abdominal movement 5 3 1 6 Sensor for body position 5 3 1 7 Sensor for leg movement 5 3 1 8 Sensor for ECG 5 3 1 9 Sensor for Snoring external 5 3 1 10 Sensors for EEG EOG EMG 6 3 2 Preparing f...

Page 4: ...thorax effort to fasten the MiniScreen Pro unit to the patient Flexible belt with integrated pressure transducer to record abdominal effort Battery charger Shoulder bag for storing the MiniScreen Pro...

Page 5: ...more than one patient at any one time is not allowed Warning When applying sensors to the patient please ensure that no electrical line of the device is in contact with any other electrical conductive...

Page 6: ...arties LM shall not be liable for damages arising from non observance of this instruction manual The actual guarantee and warranty condi tions in the LM terms of sale delivery are not exten dable 3 Re...

Page 7: ...hing of the belts may result in poor signals and possible loss of data 3 1 6 Sensor for body position The position sensor integrated in the MiniScreen Pro displays information related to the patient s...

Page 8: ...switch the unit on before going to bed by sliding the switch to the right 7 The MiniScreen Pro is now ready for use The unit can now be disconnected from the PC Note remove cable by pulling on the plu...

Page 9: ...onnection tab followed by Read measurement from device or select the following tab button 3 4 Online Measurements Inpatient Please adhere to the following steps when perfor ming an online measurement...

Page 10: ...orage are between 5 C and 45 C The maximal humidity may not exceed 95 condensation has to be avoided 5 Evaluation software for PC 5 1 Installing the software To install the MiniScreen Pro software pla...

Page 11: ...ly on the patient and remove any nail polish that may be present A small red light should be present in the sensor when the finger is inserted If the light doesn t switch on when the sensor is applied...

Page 12: ...metric measurement and light Intensity display CPAP Bi level Differential pressure measurement directly on CPAP mask Measurement range 0cmH2O 45cmH2O 5 Leg movement Two leg sensors EMG for measurement...

Page 13: ...4 Information 3 Installing the software 8 Leg movement 5 Maintenance 8 Optical isolator 3 Pacemaker 3 Position sensor 5 Preparing for measurement 6 Pressure pads 5 Safety related information 3 Selecti...

Page 14: ...10 EC Declaration of conformity 12...

Page 15: ......

Page 16: ......

Reviews: