Lowenstein Medical 200-0902 Description And Instructions For Use Download Page 10

4.  Service and Maintaining the device

4.1.  Charging the battery

The MiniScreen Pro unit is equipped with a special
fast-charge Li-Ion battery. The battery charging
adaptor included with your device is specially
designed for this type of battery and should only be
used with your MiniScreen Pro. It is recommended to
charge

the

battery

completely

after

each

measurement.

Caution: It is not allowed to use a different
battery charger than the one provided!

Battery charging procedure

 

 :

Connect the battery charger adaptor to the
MiniScreen Pro

Plug the charger into a power outlet

The charger LED will illuminate white. As soon
as the battery is charged, the LED will illuminate
blue. The device can remain connected to the
charging adaptor for a long period of time
without suffering any damage.

4.2.  Cleaning instructions

The flow prong is designed for single patient use and
should not be used more than once or on different
patients.

The finger sensor, thermistor, leg movement sensor,
the microphone and the electrodes can be cleaned
using a moist cloth or if necessary by using
disinfectant. You should never however immerse the
sensors in liquid. Any adhesive residues that may be
present should be removed regularly (e.g. using
alcohol swabs). The RLS sensor wire is very fragile.
Special care is required when cleaning this sensor.

The thorax and abdomen belts can be machine
washed at temperature of 60°C. Please ensure to
remove the pressure pads from the belts before
placing in the washing machine.

The bag can be cleaned using moist disinfectant
wipes (observe the detergent manufacturer’s instruc-
tions).

4.3.  Maintenance

We recommend the device including all accessories
is serviced every 2 years. The servicing should only
be carried out by the manufacturer or by an
authorised agent.

Calibration of the CPAP channel should be carried
out once at the first time when the device is used.

4.4.  Transport and storage conditions

The ambient temperatures for transport and storage
are between 5°C and 45°C. The maximal humidity

may not exceed 95%. condensation has to be
avoided.

5.  Evaluation software for PC

5.1.  Installing the software

To install the MiniScreen Pro software, place the CD
supplied with your device into the CD-ROM drive of
your computer. The program will automatically run. If
the program does not run automatically open the
Explorer option on your operating system and select
the CD-ROM drive. Once selected double click the
Setup.exe file from the software.

You can check your computer and network (if
applicable) for MiniScreen software versions that
may have been previously installed. To ensure the
correct up to date version is selected, open the
software installed and click the Search button. The
program will then automatically find and use the
latest version installed.

Note: Alternatively you can update the software by
selecting menu item Tools / Update from the soft-
ware. (Only for updating the software).

After selecting the target drive the installation pro-
gram will now automatically create the necessary
directories and copy all the required files to the hard
drive. The installation will automatically insert a new
group containing an entry for the new software in the
Programs folder of the Start Menu. In addition a
shortcut on your Windows desktop will be created.

After successful installation, remove the CD and
store it in a safe place.

5.2.  Selecting a printer

The measurement curves and reports can be printed
on any commonly available type of printer (ink jet,
dot-matrix or laser printer) that is already installed on
your Windows system. The correct printer driver can
be selected using the Windows Control Panel
application.

6.  Troubleshooting

Channels (e.g. CPAP) are missing in the display.

The channels are deactivated and therefore do not
appear during ”Test” and ”Record”.

Check which channels are active for recording using
the menu item Options / 

Channel settings.

Channels are missing after loading a measure-
ment.

They were not recorded or have been hidden for
measurement data display.

Flow signal is missing during recording or
follows the edge of the range.

The flow signal follows a straight flat line in the

- 8 -

Summary of Contents for 200-0902

Page 1: ......

Page 2: ......

Page 3: ...ax and abdominal movement 5 3 1 6 Sensor for body position 5 3 1 7 Sensor for leg movement 5 3 1 8 Sensor for ECG 5 3 1 9 Sensor for Snoring external 5 3 1 10 Sensors for EEG EOG EMG 6 3 2 Preparing f...

Page 4: ...thorax effort to fasten the MiniScreen Pro unit to the patient Flexible belt with integrated pressure transducer to record abdominal effort Battery charger Shoulder bag for storing the MiniScreen Pro...

Page 5: ...more than one patient at any one time is not allowed Warning When applying sensors to the patient please ensure that no electrical line of the device is in contact with any other electrical conductive...

Page 6: ...arties LM shall not be liable for damages arising from non observance of this instruction manual The actual guarantee and warranty condi tions in the LM terms of sale delivery are not exten dable 3 Re...

Page 7: ...hing of the belts may result in poor signals and possible loss of data 3 1 6 Sensor for body position The position sensor integrated in the MiniScreen Pro displays information related to the patient s...

Page 8: ...switch the unit on before going to bed by sliding the switch to the right 7 The MiniScreen Pro is now ready for use The unit can now be disconnected from the PC Note remove cable by pulling on the plu...

Page 9: ...onnection tab followed by Read measurement from device or select the following tab button 3 4 Online Measurements Inpatient Please adhere to the following steps when perfor ming an online measurement...

Page 10: ...orage are between 5 C and 45 C The maximal humidity may not exceed 95 condensation has to be avoided 5 Evaluation software for PC 5 1 Installing the software To install the MiniScreen Pro software pla...

Page 11: ...ly on the patient and remove any nail polish that may be present A small red light should be present in the sensor when the finger is inserted If the light doesn t switch on when the sensor is applied...

Page 12: ...metric measurement and light Intensity display CPAP Bi level Differential pressure measurement directly on CPAP mask Measurement range 0cmH2O 45cmH2O 5 Leg movement Two leg sensors EMG for measurement...

Page 13: ...4 Information 3 Installing the software 8 Leg movement 5 Maintenance 8 Optical isolator 3 Pacemaker 3 Position sensor 5 Preparing for measurement 6 Pressure pads 5 Safety related information 3 Selecti...

Page 14: ...10 EC Declaration of conformity 12...

Page 15: ......

Page 16: ......

Reviews: