19
12. Technical data
Parameter
Indicating ranges
Resolution
Conductivity
0.0 ... 200.0 µS/cm
0.1 µS/cm
0 ... 2000 µS/cm
1 µS/cm
0.00 ... 20.00 mS/cm 0.01 mS/cm
0.0 ... 200.0 mS/cm
0.1mS/cm
Specific Resistivity 0.005 ... 100.0 kOhm 0.001 kOhm · cm,
0.01 kOhm · cm
0.1 kOhm · cm
Total dissolved
0 ... 1999 mg/l
1 mg/l
solids (TDS)
Salinity
0.0 ... 70.0
0.1
Temperature
-5.0 ... +100.0°C
0.1°C
23.0 ... 212°F
bzw. 0.1°F
• Accuracy:
conductivity, spec. resistance, TDS and (at nominal
temperature) salinity (±3 digits) ±0.5% of result±0.5%
FS
temperature ±0.3K
• Cell correction:
determination with automatic
calibration in the range
from 0.55 to ± 0.11 1/cm
• Temperature
choice of
compensation:
- linear temperature compensation with
a compensation factor of 0.3 to 3.0 %/K
- non-linear temperature compensa-
tion in acc. with EN 27888 (DIN 38404)
reference temperatures 20°C and 25°C
or without temperature compensation
Summary of Contents for SensoDirect Con200