6.3.2. Rotor, seal and filter maintenance
Warning!
Do not forget any objects or tools inside the air filter.
DARWIN M
•
Release and remove the rear lid.
•
Extract
the
FPS
filter
(
for
D 3000: FPSE,
the net and
FPSI)
and the net from the rear lid.
•
Remove all deposits (metal, grease, etc.) from the drainage ring of the rear lid
and from the air filter
shell (see 6.3.4).
•
Replace
seals
GSM
and
GTI
if necessary
•
Clean or replace the
FPS (
for
D 3000: FPSE
and
FPSI)
filter (see 6.3.3).
•
Insert
the
FPS (
for
D 3000:
insert
FPSI,
the net and
FPSE)
filter on the net of the rear lid.
•
Reassemble and secure the rear lid.
•
Remove the two safety pins mounted on hooks of the front lid.
•
Release and remove the front lid.
•
Extract
filter
FCSE (FCS
for
D 600)
from the inside of the centrifuge.
•
Clean or replace the
FCSE (FCS
for
D 600)
filter (see 6.3.3).
•
Remove all deposits from the surface of the centrifuge (metallic dust, grease, etc. see 6.3.4.)
•
Reposition
filter
FCSE (FCS
for
D 600)
inside the centrifuge
•
Replace
the
GTI
seal if necessary
•
Reassemble and secure the front lid inserting the two safety pins into the hooks.
DARWIN D
•
Release and remove the rear lid.
•
Extract
the
FPS
filter
(
for
D 3000: FPSE,
the net and
FPSI)
and the net from the rear lid.
•
Remove all deposits (metal, grease, etc.) from the drainage ring of the rear lid
and from the air filter
shell (see 6.3.4).
•
Replace
seals
GSM
and
GTI
if necessary
•
Clean or replace the
FPS (
for
D 3000: FPSE
and
FPSI)
filter (see 6.3.3).
•
Insert
the
FPS (
for
D 3000:
insert
FPSI
the net and
FPSE)
filter on the net of the rear lid.
•
Reassemble and secure the rear lid.
•
Remove the two safety pins mounted on hooks of the front lid.
•
Release and remove the front lid.
•
Extract
filter
FCSE (FCS
for
D 600)
and
FCSI
from the inside of the centrifuge.
•
Clean or replace the
FCSE (FCS
for
D 600)
filter (see 6.3.3).
•
Remove all deposits from the surface of the centrifuge (metallic dust, grease, etc. see 6.3.4.)
•
Reposition
filter
FCSE (FCS
for
D 600)
and
FCSI
inside the centrifuge
•
Replace
the
GTI
seal if necessary
•
Reassemble and secure the front lid inserting the two safety pins into the hooks.
DARWIN T
•
Release and remove the rear lid.
•
Extract
the
FPS
filter
(
for
D 3000: FPSE,
the net and
FPSI)
and the netfrom the rear lid.
•
Remove all deposits (metal, grease, etc.) from the drainage ring of the rear lid
and from the air filter
shell (see 6.3.4).
•
Replace
seals
GSM
and
GTI
if necessary
•
Clean or replace the
FPS (
for
D 3000: FPSE
and
FPSI)
filter (see 6.3.3).
•
Insert
the
FPS (
for
D 3000:
insert
FPSI,
the net and
FPSE)
filter on the net of the rear lid.
•
Reassemble and secure the rear lid.
•
Remove the two safety pins mounted on hooks of the front lid.
•
Release and remove the front lid.
•
Remove all deposits from the surface of the centrifuge (metallic dust, grease, etc. see 6.3.4.)
•
Replace
the
GTI
seal if necessary
•
Reassemble and secure the front lid inserting the two safety pins into the hooks.
EN