Introduction

TS series products are controllers of four channels and multi-function
syringe pumps. They need to be connected with drive units to constitute
syringe pump systems of strong functions and easy-to-operate. They are
widely used in biological laboratory. There are 3 types:

Controller Type

T S - 1 A

T S - 2 A

T S - 1 B

Drive Unit Type

L 0 1 0 7 - 1 A

L 0 1 0 7 - A

W 0 1 0 9 - 1 B

2

Working Mode

Infusion-only

Infusion/Withdrawal

Infusion/Withdrawal

S1 Controller

S2 Drive Unit

S3 Glass Syringe

Note:

Controller and drive unit are connected through 8 pin S terminal.
Be careful when insertion or extraction. Please see page 21 for
detail information about

control function.

External

F1 Power Switch

F2 Power Socket

F3 Drive Unit Socket

F4 External Control Input and Communication Interface

F5 External Control Output Interface

Operation Panel

SYRINGE PUMP OPERATING MANUAL

SYRINGE PUMP OPERATING MANUAL

Fast Forward Key

Exit Key

Channels Switch Key

Fast Reverse Key

Start/Stop Key

Rotary Coded Switch

Basic Operation

Start/Stop Key
Press the Start/Stop Key to start or stop the pump. In multi-
channel working mode, to pause the running state, the pump will
prompt if continue the current working state.
Fast Forward Key
In multi-channel working mode, pressing the Fast Forward Key all
the enabling channels infuse at the max. speed. In single channel
working mode, pressing the Fast Forward Key this channel infuses
at the max. speed. Loose this key the pump stops. Other keys are
invalid when pressing this key.
Fast Reverse Key
In multi-channel working mode, pressing the Fast Reverse Key all
the enabling channels withdraw at the max. speed. In single
channel working mode, pressing the Fast Reverse Key this channel
withdraws at the max. speed. Loose this key the pump stops. Other
keys are invalid when pressing this key.

Note: Fast Forward Key and Fast Reverse Key are invalid when the
pump is in running state.

Rotary Coded Switch (RCS)
Turn the Rotary Coded Switch for menu selection or parameter
setting. Press the Rotary Coded Switch for confirmation.

Note: Put the finger in the center of the switch to press the RCS.
Put the finger on the small concave which is on the side of the
switch to turn the RCS.

Channels Switch Key
Switch between different channels.

Exit Key
Cancel current operation and return to previous menu.

Summary of Contents for TS-1A/L0107-1A

Page 1: ...ease read operating manual carefully before operation Over push or over draw the syringe may result in the fluid sprayed Use appropriate measures to protect operator and equipment Be careful during op...

Page 2: ...Key to start or stop the pump In multi channel working mode to pause the running state the pump will prompt if continue the current working state Fast Forward Key In multi channel working mode pressi...

Page 3: ...me 60sec Act 50 00 L Display Select Volume Speed Flow Multi channel Running Interface Switch on the pump First initializing and then display the running interface a Enabling symbol Symbol of enabling...

Page 4: ...ect if save the change Set Delay Time Set delay time to start each drive unit in turn Time units are sec min and hour Time numerical value is from 0 1 to 999 9 The setting method is as below Turn the...

Page 5: ...unning Interface SYRINGE PUMP OPERATING MANUAL SYRINGE PUMP OPERATING MANUAL Press the RCS Press the Exit Key Press the Exit Key Press the RCS Press the Exit Key Press the RCS Press the RCS Select Yes...

Page 6: ...the Syringe If the syringe used is not listed in the menu the inner diameter of the syringe needs to be measured and entered 0 01 mm 40 mm The pump can save 4 inner diameters of the syringes Please se...

Page 7: ...ing Select No not to save the setting Select Cancel to return to previous menu and keep on selecting Note When the one time dispensing volume or the total dispensing volume multiply one time dispensin...

Page 8: ...selecting Turn the to select if save the parameter RCS Note When the pump can t withdrawal the setting volume in the setting time the running interface will display W speed Overflow see below Please...

Page 9: ...tion volume There are 10 50 and 80 of the syringe volume available When selecting the user defined syringe the default calibration volume is 50 of the syringe volume Please see page 9 Syringe Select t...

Page 10: ...cation Flow Rate L min Glass Syringe SYRINGE PUMP OPERATING MANUAL SYRINGE PUMP OPERATING MANUAL 10 25 50 100 250 500 1000 1 318nL 13 18 L 3 320nL 33 20 L 6 610nL 66 10 L 13 282nL 132 82 L 32 96nL 329...

Page 11: ...htening knobs according to t h e a r r o w h e a d directions to fix the glass syringe 2 See above drawing to load the glass syringe Lift up the p r e s s k n o b a n d rotate the press knob according...

Page 12: ...f them Delaying Startup Delaying startup time of each channel can be controlled separately Memory Function Select resume operation or remain stopped when power returns after an interruption Block Prot...

Page 13: ...ight 3 Kg IP Rating IP41 23 24 TS 1A L0107 1A TS 2A L0107 2A Specifications TS 1B W0109 1B Specifications Max Infusion Distance 70 mm Acceptable Glass Syringe 5 L 1000 L Linear Speed 7 94 m min 79 4 m...

Page 14: ...ecision Pump Co Ltd SYRINGE PUMP OPERATING MANUAL Baoding Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051...

Reviews: