Cabling should be screened and routed away from noise-carrying cables and
machines to reduce interference. The screen can be connected to power earth if
it is not already grounded at the shunt end.
The shunt should be located in the low voltage unswitched return line to
minimise common-mode noise.
These connectors are only used when options are fitted
Connections
E - +
Common
No function
Disp. Current
Reset total
IMPORTANT!
Set the right-hand switch to its upper ‘LOCK’ position after setup, to save your settings.