Inspection and Cleaning
Document Number: 814212-2.0-ENG
Original Document
Yearly
1.
Check the detector inserts for wear.
2.
Check all connectors and terminals within the electrical services cabinet.
3.
Confirm that the 12V and 24V DC supplies are within the 5% limits.
4.
Check the operation of all cooling fans.
5.
As a minimum the x-ray oil should be changed annually, unless the machine specific
details state otherwise. The oil is of an insulating type specific for x-rays and any air
bubbles must be purged before generating x-rays. Loma Service office can perform
the oil change on request. The customer will be responsible for the environmental
disposal of the old oil. For more information on Maintenance see the reference guide
Replacement oil should only be obtained from Loma Systems as standard
oil direct from the manufacturer requires additional processing to meet the
specification required.
6.
For more detailed information on Maintenance see the relevant X
5
Reference Guide.