X-Ray Set
The X-Ray Set tab of the
screen is the default tab displayed when the screen
opens. It provides current status information for a number of system areas and these are
listed and described below.
X-Ray Set - This area contains a coloured status panel for each system area that is
being monitored:
Green
- A green coloured panel indicates that the interlock is active and
working correctly.
Red
- A red coloured panel indicates a problem with an interlock which must
be rectified before the system can be restarted and X-Ray generation
resumed.
Grey - A grey coloured panel indicates that the interlock is inactive which
does not indicate a fault.
System Area - Buttons are provided for each area, as listed and described below,
which when selected will then display applicable information in the screen.
The system includes a number of lockable panels that provide access to
various parts of the system to carry out cleaning, maintenance and servicing
operations.
To ensure that users are not accidentally exposed to X-Rays each of these
access panels is fitted with a switch which stops the system conveyor and the
generation of X-Ray s automatically if an access panel is opened. The status
of each panel is monitored and can be reviewed at any time. This section
provides further details.
The system monitors the status of a number of key items to ensure that it is
operating correctly. More details are available in this section.
The status of a number of key X-Ray parameters is constantly monitored by
the system and warnings are displayed if their values exceed defined limits. A
test may also be carried out to confirm that X-Ray s are being generated at
the correct power level. This section includes full details.
The temperature of key system components is constantly monitored to ensure
that they stay within defined limits.
All signals utilised by the system to ensure that the reject device and
associated PECs are operating correctly can be tested at any time. This
section includes further details.
In addition, the following buttons are displayed when the X-Ray Set tab is selected:
Clean - Tapping on this button puts the system temporarily into clean down mode for 20
seconds and displays a progress bar which counts down the time.
Reset Faults - Once any faults displayed by the system have been resolved, tapping on this
button resets the status of the applicable item.
OK - Tapping on this button closes the
screen. The