Operating instructions
ScanAfia X TK, S, ICU Version 1.5
Copyright © Lojer Oy, 2015
26
6.9
Equipotential connection
The purpose of the equipotential is to prevent the occurrence of dangerous
voltage differences between conductive components that are touched
simultaneously. The ScanAfia X hospital beds have the equipotential
connection in the front element below the equipotential connector label. Use the
equipotential whenever it is needed; note the manufacturer's operating
instructions. Attach the equipotential cable to the connector with an appropriate
attachment.
Figure 56.
Equipotential connector label
Figure 57.
Equipotential connector at the end of the bed
IMPORTANT!
Always disconnect the equipotential cable before moving the
bed.
6.10
Operating the battery of the hospital bed
The battery of the hospital bed is intended for use in emergency situations only,
in which no mains power is available, such as during power outages and when
moving the bed.
IMPORTANT!
The battery sounds a signal and stops the movement for a
moment when its power is low. If the battery power is low and
bed movements must be run, run them one movement at a
time. Do not use the pushbuttons of the hand-held control,
which activate several movements simultaneously.