background image

6

 

 

Copyright © Lojer Oy, 2013 

02/2013 Rev. 1.0 

2.2

 

Before use 

The device 

is intended to be used in normal, dry indoor conditions. Ensure that the temperature of the room is b10…40 

°C and the humidity is within the range of 30…75 %. If there is chance that device has been exposed to temperatures below 0°C

allow it to adjust to the indoor temperature for at least 5 hours before using any of its features. 

Familiarise yourself with the instructions and carry out the following before using the table: 

 

Make sure that all packing materials have been removed  

 

Make sure that the device can move freely uo and down. 

 

Place the device in the location where it will be used and lock the castors. 

 

Connect the power plug to a socket whose supply voltage corresponds to the voltage shown on the device’s 

type plate. Make sure that the cord runs freely from the connection box.  

 

For safety reasons always connect the power cord to grounded socket. 

 

Do not bind the power cord to the device as the lifting motion can severe the cord. Ensure that the cord is easily 
detachable in an emergency situation. 

 

Make sure that the distance to the socket is not more than 2 meters. 

 

Always detach 

the power cord before moving the device.  Make sure that the cord doesn’t  get  stuck between 

parts of the frame or under the castors. 

 

If the power cord is damaged, unplug it immediately. Do not use the device and contact the service. Use only the 
original power cord. 

 

Make sure that the patient doesn’t accidentally move/touch the place free bar or any other control device.

 

 

Do not place anything under the device (e.g. chair, because the device has to be able move freely). Make sure 

that the patient’s limbs do 

not get caught in the frame of the device. 

 

Do not place the device under any wall structures or too close to the wall. 

 

Do not modify the structure of the device or install parts other than those mentioned in this document. 

 

Do not use the device or the acce

ssory if it doesn’t work properly. 

Contact the service 

 

Do not lift the device from the head section, arm rests or the side supports. Do not push the device on to a door 
sill. 

 

 

WARNING! 

Children or people with no experience of the device or those with restricted understanding must not 

use the device. Children must be supervised to ensure that they do not play with the device! For safety reasons 
use the safety switch or unplug the power cord when the device is left unsupervised. 

3

 

Using the treatment table 

Note! 

Do not use the electrical functions of the device non-stop for longer than the permissible two (2) minutes. Longer continuous use may 

cause the transformer to overheat. If you use electrical functions non-stop for two (2) minutes, keep to the operating time ratio and do not use 
any electrical functions for 18 minutes. 

3.1

 

Maximum inclination 

The height adjustment range of the Delta treatment tables: 

Standard; 1MD4 57-87 cm, 1MD7L and 2MD602 58-89 cm 

Professional; 1MD4 47-98 cm, 1MD7L and 2MD602 48-100 cm. 

The foot section adjustment range to 75 ° angle.  

Summary of Contents for 1MD1X X-frame

Page 1: ...Lojer Delta treatment tables Instructions of use 18 2 2013 ...

Page 2: ...ng the middle section drainage 8 3 5 Adjusting the foot section 8 3 6 Adjusting the head section 8 3 7 Adjusting the arm rests 9 3 8 Adjusting the side supports 9 3 9 Double hinged face section accessory 9 4 Cleaning and disinfecting 10 4 1 Metal and plastic surfaces 10 4 2 Upholstery 10 5 Maintenance 11 5 1 Biannual measures 11 5 2 Annual measures 12 5 3 Troubleshooting 12 5 4 Preventive maintanc...

Page 3: ... instructions for operating and maintaining Lojer treatment tables Please familiarise yourself with these instructions before using the table Use the device only as described and for the specified applications Store these instructions in an appropriate way making sure that the instructions are available to all possible users throughout the life of the device To avoid injury follow the instructions...

Page 4: ...t adjustment bar Castors Double hinged head section Handswitch Foot pedal Frame for foot pedal Standard and Professional 1 Castors 2 Place free height adjustment bar 3 Arm rest adjustment lever 4 Arm rest 5 Head section head section adjustment lever 6 Middle section 7 Foot section 8 Side support 9 Central locking pedal 1 Castors 2 Head section head section adjustment lever 3 Lying surface 4 Centra...

Page 5: ...le isolated Warning labels squeezing hazard placed on the place free height adjustment bar or on the lower frame 2 Introduction 2 1 Inspection upon delivery Before the device is taken into use check that the packaging is intact and that it has not been damaged during transportation Please notify the transport company and the supplier of any transit damage within two 2 days of receiving the deliver...

Page 6: ...rvice Use only the original power cord Make sure that the patient doesn t accidentally move touch the place free bar or any other control device Do not place anything under the device e g chair because the device has to be able move freely Make sure that the patient s limbs do not get caught in the frame of the device Do not place the device under any wall structures or too close to the wall Do no...

Page 7: ...t adjustment The height of the table can be adjusted with pneumatic foot control Figure 5a Height can be adjusted also with place free height adjustment bar accessory Figure 5b Make sure that under no circumstances the patient is not able to touch the foot control or the place free bar Use the safety switch when necessary a b SQUEEZING HAZARD Make sure that nothing is between the structure or unde...

Page 8: ...tween the structure or under the device during lifting lowering 3 5 Adjusting the foot section Make sure that the middle section is locked before adjusting the foot section Lift the foot section to desired position It will lock to its place automatically Before lowering free the locking and push the section down SQUEEZING HAZARD Make sure that nothing is between the structure or under the device d...

Page 9: ...est Max load 30 kg Do not lift the table on the arm rests 3 8 Adjusting the side supports Lift the side support up and push it towards the head section in order to lock it Figure 9 In order to lower the support push it towards the foot section and turn it down Do not sit on the side support Maximum load on the side support is 30 kg Do not lift the table from the side supports 3 9 Double hinged fac...

Page 10: ... surfaces may get damaged if aromatic hydrocarbons benzene and its derivates ketones ethers esters and chlorinated hydrocarbons are used Plastic might also deteriorate if it is exposed to various chemicals at the same time Stainless steel surfaces are highly resistant to chemicals Use for mild detergent solution for cleaning Ammonia and most of the solvents can be used to remove difficult stains A...

Page 11: ...be held responsible for the effects of other solutions than those mentioned in the matrix Wipe the surface with a clean damp cloth after disinfecting Dried substances or substances that have penetrated for an extended period may not be able to be removed completely 5 Maintenance Always unplug the power cord before service Only trained person may carry out service and repair Maintenance carried out...

Page 12: ...g maintenance inspection and servicing to assess the safety of the devices Tests should be performed by a qualified personnel Qualification should include training knowledge and experience with the relevant test procedures technologies and regulations The personnel assessing the safety should be able to recognize possible consequences and risks related to non conforming devices Indication Defect A...

Page 13: ...nt under test change points if necessary Attach the applied parts to the measuring device Note In Class I equipment a leakage current measurement can be performed only after the protective earth testing has been passed Use the correct measurement method and procedures related to that Currents to be measured Equipment leakage current current from the mains part to earth through protective conductor...

Page 14: ...ofessional Ingress protection rating IP55 Electric classification Class II B tyypin applied part Safe Working Load 200kg Width 55 cm Length 195 cm Weight 60 75 kg Transport temperature 5 60 C humidity 30 75 Storage temperature 5 60 C humidity 30 75 Operating temperature 10 40 C humidity 30 75 6 1 Circuit diagrams Circuit diagram is shown below Figure 12 Figure 12 Circuit diagram of Delta tables Fi...

Page 15: ...the directive 7 Recycling Most of the materials used in the device are recyclable When the device is removed from usage it should be dissembled and recycled appropriately Recycling should be done by specialized company Do not dispose the device in the household waste Remove the battery from the device The oil from the hydraulic system should be removed and disposed appropriately Gas springs should...

Page 16: ... Fax 35810 830 6702 Email firstname lastname lojer com info lojer com www lojer com Your local Lojer dealer see www lojer com distributors Model ________________________________________________________ Serial number __________________________________________________ Date of purchase ________________________________________________ Your local Lojer dealer ___________________________________________...

Reviews: