LOGAN FRIABILITY TESTING SYSTEM
FAB-2S FRIABILITY TESTER
Page
10
of
12
To set speed (from 25 RPM to 70 RMP):
A.
Press and hold “Start/Stop” button and turn on the main power switch at the same time; this
will put the FAB-2S into speed program mode. In speed program mode, only the last 2 digits can
be programmed (25 RMP – 70 RMP).
B.
Press the “Set” button, the last digit will start flashing.
C.
Press “Enter” to increase the flashing digit. (from 0 to 9)
D.
Press “Set” button, and the fourth digit will start flashing.
E.
Press “Enter” to increase the flashing digit. (from 2 to 7.)
F.
Press “Reset” to enter the new speed setting into the FAB-2S.
G.
Turn off the FAB-2S and power on again to check the new speed setting on the display to start
the test at the new speed.