4
Specifications
4.1
Technical Specifications
Principle
:
eddy current brake small disk.
Mains voltage
:
115-230 VAC (adjustable)
Frequency
:
50 Hz, 60 Hz.
Patient capacity
:
350 lb - 160 kg.
Power consumption
:
120 VA.
Workload
:
0-999 W HYP, 0-999 W LIN
Device control
:
manual, external analog, external digital
Accuracy
:
< 3 W for P <= 100 W, < 3% for 100 W<
P<500W, <5% for P>500 W.
Read-out workload
:
digital, 0-999 resolution 1 W.
Read-out pedalling speed
:
analog 0-180 rpm or digital 0-255.
Output workload
:
100, 200 or 500 W 1 V (max. 999 W).
Output pedalling speed
:
100 min
-1
1 V.
Input voltage for external control
:
100,200 or 500 W 1 V (max. 999 W).
Safety class Corival
:
I type B, IEC 601-1 chapter 18, 19 and 20.
Safety class C BP option
:
IIa type B, IEC 601-1 chapter 18, 19 and 20.
Leakage current
:
< 0.4 mA.
Isolation resistance
:
> 4 M
Ω
.
Earthway resistance
:
< 0.2
Ω
.
Zero load
:
< 8 W at 60 rpm.
Noise level
:
+/- 60dB at workload 200W/60rpm +/-80cm from
crank.
4.2
Fuses
115 V
:
2 x 1,60 A slow.
230 V
:
2 x 0,80 A slow.
Met opmaak:
opsommingstekens en
nummering