S
ENSOR
C
OMMANDS AND
R
ESPONSES
42
Copyright LMI Technologies Inc. 2003
Using the high speed polling mode the data transfer can maintain the highest
sample rate possible (667 Hz) if the baud rate has been changed to 57600 baud.
4.5.2
C
OMMAND
35
P
UT
S
ENSOR
I
N
H
IGH
S
PEED
M
ODE
Purpose
Puts the sensor in high speed polling mode and changes the data
transfer protocol.
Command
[command]
command
(1 byte)
35
Response Format:
if using CRC : [Command]
Command
(Byte)
35
Success
0
Fail
If using ChkSum : None
NOTE
: The sensor will stay in the high speed polling mode even if the
sensor gets powered down and up again unless the
F
character has
been sent.