![Livoo SL250 User Manual Download Page 14](http://html.mh-extra.com/html/livoo/sl250/sl250_user-manual_1931722014.webp)
www.livoo.fr
Press the "ALARM" button for 2 seconds to enter the alarm setting mode. In the setting mode, use "
▲
" and
"
▼
" to set the required time and then the "ALARM" button to confirm the time and move to the minute
setting. Use "
▲
" and "
▼
" to set the minutes and "ALARM" to validate and exit the setting mode.
To activate the alarm, use the "ALARM" switch on the back of the station (button B9 figure 2).
If the switch is set to "ON", the icon appears next to the current date display (screen A1 figure 1) and the
alarm is activated. If the switch is set to "OFF", the icon will not appear and the alarm will be disabled.
The alarm sounds for 2 minutes unless you turn it off by pressing any button. If the alarm is turned off, it
will only automatically repeat after 24 hours.
If you press the "SNOOZE/LIGHT" button (button B2, Figure 2) while the alarm is sounding, it will not be
disabled. It will go off 5 minutes after you press this button.
The alarm will sound louder and louder during the 2 minutes and change its volume 4 times.
Outdoor sensor
To power up the outdoor sensor, insert 2 AA batteries, observing the correct polarity (Figure 3). When you
insert the batteries, a brief red light will appear indicating that the sensor is powered on. The sensor will
automatically transmit the information to the weather station.
Briefly press the "
▼
" button to select a communication channel. The weather station has 3 channels
(from 1 to 3). The sensor included with the unit transmits the temperature, humidity and air pressure
trend of its surroundings on channel 1 of the station.
If the station is unable to receive the information transmitted by the outdoor probe, the station displays
'--' as the value.
To restart the connection between the station and the probe, hold down the "
▼
" button for 3 seconds.
Maximum and minimum
By pressing the "MEM" button (button B1 in Fig. 2), the instrument will display the maximum temperature
and humidity detected by your station since it was last switched on, both those detected by the station
and those detected by the probe via channel 1.
Pressing the "MEM" button again will cause the station to display the minimum values.
The display is symbolized by the appearance of the "MAX" or "MIN" logo on screens A5 and A6.
To reset the stored maximum and minimum values, press and hold the MEM button for about 2 seconds.
Outdoor Temperature Alert
Briefly press the ALERT button to activate the temperature alert function.
If the alert is activated, the "TEMP ALERTS" symbol will appear next to the outdoor temperature (screen A5
figure 1). The alarm signal will sound when the outdoor temperature falls below the set minimum alert
temperature or rises above the set maximum alert temperature.
To set the maximum and minimum alert temperatures, press and hold the ALERT button for 2 seconds.
The outdoor temperature will flash and two icons will appear "Hi" and "Lo" (screen A5 figure 1). When the
"Hi" icon is flashing, you can set the maximum alert temperature. By briefly pressing the "ALERT" button,