33
9.10
Accessories
The use of incompatible accessories can injure the patient!
Use original accessories from the manufacturer only.
Warning
NOTE
The manufacturer is not responsible for the use of unapproved accessories.
9.10.1 Lifting Pole
Fig.
Lifting Pole
To ensure safe use of the lifting pole:
Never exceed the maximum load of 75 kg.
Never use the lifting pole for rehabilitation exercises.
To prevent the bed from tipping over, ensure that the lifting pole does not project out from the bed.
Replace the plastic handle every 4 years.
Lifting Pole Positions:
Over the backrest (working position).
Parallel to the head board.
To install the lifting pole:
Insert the lifting pole in the corresponding sleeve fittings at the head of the bed (corners).
Ensure that the safety pin locks into place.
A plastic grab handle with an adjustable strap shall be attached to the lifting pole.
NOTE
The date of manufacture is marked on the grab handle. LINET
®
recommends that you replace the
plastic grab handle every four years.