3. Program the Cell Monitoring Voltage Thresholds

The essential function of the LTC6811 is to measure and 

report the voltage on each battery cell when commanded. 

First click on the STARTCELL button. This commands an 

A/D conversion of all 12-cell voltages in the bit-resolution 

configured from the selected CONV MODE (Conversion 

Mode) in the ADC SETTINGS box. The actual cell voltage 

measurements are displayed in groups of 3 when each 

command button inside the READ CELLS box is clicked. 
 

NOTE:

 The READ DISPLAY box has options to display 

Cell Voltages in VOLTS or HEXadecimal. First select 

VOLTS or HEX then click on one of the command but-

tons inside READ CELLS box. Every time a command 

button inside the READ CELLS box is clicked; new data 

is downloaded from the board and displayed as selected 

in VOLTS or HEX.

5. Discharge Cells

OPERATING THE CONTROL SCREEN

In the section labeled SET VOLTAGE LIMITS, click on the 

boxes and enter voltage values for the overvoltage and 

undervoltage thresholds appropriate for the cells being 

monitored. The voltage value entered will be rounded to 

the actual value used by the LTC6811 and displayed in 

the box. The voltage ranges for these thresholds is 0V to 

6.5520V.
These monitor thresholds can be applied globally to each 

and every cell in the system or customized for the cells 

connected to an individual board by clicking the desired 

option button. Individual boards are selected for program-

ming by the left hand tabs in multiple board systems.

4. Read Cell Voltages

The LTC6811 includes the ability to remove energy from 

individual cells to help equalize the cell charges within a 

stack of batteries. DC2259A contains a P-channel MOSFET 

in series with a 33Ω resistor across each cell connection. 

When enabled, a cell is loaded and charge is pulled from 

the cell with energy dissipated in the switch and resistor.
A check box is provided for each cell to be discharged. 

Checking this box (Cell 6 in the above example screen 

shot) and then writing the new configuration with a WRITE 

CONFIG button click will load the cell.
The discharge transistors are automatically turned off 

momentarily while the A/D converter is measuring the 

cell voltage using the normal STARTCELL command. This 

Summary of Contents for LTC6811-1

Page 1: ...ogy and the Linear logo and Linduino are registered trademarks and QuikEval is a trademark of Linear Technology Corporation All other trademarks are the property of their respective owners Photo 1 Connecting a Two Board Stack to a PC Design files for this circuit board are available at http www linear com demo DC2259A Alternatively for a simple demo connect twelve 100Ω resistorsbetweeneachcontactf...

Page 2: ...2 dc2259af DEMO MANUAL DC2259A Photo 2 Connections and Jumpers DESCRIPTION ...

Page 3: ...3 dc2259af DEMO MANUAL DC2259A Figure 1 Simple Cell Simulator Using Resistors DESCRIPTION ...

Page 4: ...sed n Reconfiguration of the first DC2259A to have ISOMD jumpers JP1 JP2 and JP3 changed to the 1 position like the other boards SOFTWARE SETUP To use the DC2026 USB interface board download the free program called QuikEval from http www linear com designtools software TheGUIprogramisprovidedinazippedfile LTC6804 6811_ GUI_Vxx zip 1 Unzip this folder 2 Run the setup exe file to start installation ...

Page 5: ...should be a match green PEC oval on stack display When any configuration information is changed on the screen the WRITE CONFIG command button will be il luminated serving as a reminder that this command still needs to be executed Set REFUP to 1 followed by a WRITE CONFIG command The Watchdog Timer WDT will now be prevented from expiring by issuing a repeating POLL ADC command and the stack is read...

Page 6: ...sholds appropriate for the cells being monitored The voltage value entered will be rounded to the actual value used by the LTC6811 and displayed in the box The voltage ranges for these thresholds is 0V to 6 5520V These monitor thresholds can be applied globally to each and every cell in the system or customized for the cells connected to an individual board by clicking the desired optionbutton Ind...

Page 7: ... case 7 Continuous Operation OPERATING THE CONTROL SCREEN For convenience the control panel allows for continuous operation of the DC2259A board When the command button labeled START CONTINUOUS READ CELLS is se lected the GUI operates in a continuous loop executing the following command sequence automatically n Start cell voltage n Read cell voltage groups A 1 3 B 4 6 C 7 9 and D 10 12 n Read flag...

Page 8: ...Monitor LTC6804 2 Click on the START CONTINUOUS READ CELLS com mand button to start datalogging onto a new CSV file ElapsedTime TotaltimeofDatalogfromstarttofinish Total Data Points Total data points of Datalog from start to finish Refresh Time in seconds S between each Read Cells command also known as Sample or Update Rate 3 ClickontheSTOPCONTINUOUSREADCELLScommand button to stop datalogging and ...

Page 9: ...M and set the Ref Up and voltages accordingly COLOR CODED STATUS PANEL The color coded status panel will expand to include all boards connected in a stack Each small square in this array represents an individual battery in the stack of boards The intent of this display is to provide a way to see the status of all cells at a glance The significance of the colors used is explained in the legend on t...

Page 10: ...TORY INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE EXCEPT TO THE EXTENT OF THIS INDEMNITY NEITHER PARTY SHALL BE LIABLE TO THE OTHER FOR ANY INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES The user assumes all responsibility and liability for proper and safe handling of the goods Further the user releases LTC from all claims arising from the handling or use o...

Page 11: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Analog Devices Inc DC2259A ...

Reviews: