Page 16
Display Functions and Indications
Display Functions and Indications
Functions
Timer display
The time display allows a woman to know how many minutes she
has been pumping. The pump timer starts timing when the pump
cycling starts. The time display shows the number of minutes of
pumping. It’s recommended that you only pump for 15-20 minutes
or for 2 minutes after milk stops flowing. The pump will not
automatically turn off. It’s recommended that you turn off the pump
first before removing breast cups from breast and place storage
containers in storage holders to prevent milk flowing into tubes. The
timer also displays the vacuum and cycle speed after 3 seconds from
when the setting was changed.
Problem indication
The pump is designed to check for fault conditions. If a fault is
detected, the pump will stop and a group of 3 flashing lights on the
minute timer (3 and 4) will indicate the problem. Only unplugging
the pump will clear the message and allow the pump to reset (not
applicable for Flashing 4,5,6, lights).
Flashing 1, 2, 3 Lights
This fault means that the temperature inside the pump enclosure is
higher than expected. After the pump cools, it may be used again.
This is not typical, so if it happens repeatedly contact Limerick as the
pump may need to be returned for repair.
Summary of Contents for PJ's Comfort
Page 1: ...Owners Manual ...