Page 14
Controls and Features
Note:
When vacuum strength is increased, the cycle speed will
decrease slightly.
Secondary Features
Dry Mode:
To enter
Dry Mode
, turn the pump on and hold down the
cycle speed/power knob (1) until the number 5 light turns on (this will
take about 5 [five] seconds). Release knob and the light will begin to
blink rapidly. Attach filter and tubing without breast cup assembly
and press power to start drying mode. Pump will turn on for 5 (five)
minutes and display remaining time. If vacuum level gets too high
the pump will stop and the number 5 light will begin to blink rapidly
again. To cancel drying, press cycle speed/power knob while pump is
running and pump will turn off and reboot.
Find your comfort settings
Recommended settings for first-time use
Cycle speed of 30 (thirty) and a vacuum setting of 1 (one).
The pump will automatically start up with these settings after being
unplugged or if the pump has been off for more than 5 minutes.
Adjusting vacuum strength and cycle speed
Vacuum:
Pump for the first minute or two before adjusting
the vacuum. It’s recommended to increase the vacuum in
small increments up to the level that feels comfortable to you.
Recommended pumping time is about 15-20 minutes or 2 minutes
after milk stops to flow.
Cycle:
The cycle speed knob is used to fine-tune the suction rhythm
to best mimic your baby’s suckling. It’s recommended that you
wait to adjust the cycle speed until you’ve pumped several sessions
Summary of Contents for PJ's Comfort
Page 1: ...Owners Manual ...