LMS6002D Quick Starter Manual for Evaluation Board
79 |
P a g e
© Copyright Lime Microsystems
Rev: 2.2
Last modified: 03/05/2012
Select File – Open Project – then select the project that was previously saved from the window
file box.
When loading files the software GUI tool can be setup to either load variables into the tool, but
not download them to the chip or to download the variable into the tool and auto download them.
Auto Download is enabled by the Options menu as shown below.
Figure 82 Auto Download feature
If Auto Download is not enabled then the whole SPI map can be sent to the chip by using the
download button
as shown below.
Figure 83 Download Button for Previously Saved Setup
Note
: The Ref clock frequency is not strictly part of the SPI information, it is used to calculate
the divider ratios for the PLLs. When you start the software for the first time the default clock
frequency is 40MHz. You need to change this to the TCXO frequency which typically is 30.72
MHz. If you use a different TCXO frequency you need to change the reference clock (Tools-
>Reference Clock).